Z-0145-2024Class I

Monoject 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack

Medtronic, Inc. / initiated 2023-09-20 / Open, Classified / root cause: supplier

Cardinal Health 200, LLC began a recall of Monoject 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack on . FDA classified it as Class I and gives the reason as "Demonstrated recognition and compatibility issues with syringe infusion pumps". As of , the recall status on record is Open, Classified, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-0145-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Monoject 1 mL Tuberculin Syringe Luer-Lock Tip Soft Pack
Product code
FMF Syringe, Piston
Reason for recall
Demonstrated recognition and compatibility issues with syringe infusion pumps.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
Cardinal Health notified customers on about 09/20/2023 via "URGENT MEDICAL DEVICE PRODUCT CORRECTION" letter sent FedEx next day delivery. Customers were instructed to: 1. REVIEW your inventory for the affected product codes and lots. Location of product code and lot are shown in above listed table and below example labeling. 2. COMMUNICATE with all personnel that utilize the Cardinal Health Monoject" Luer-Lock Tip syringes (1, 6, 12, 20, 35 and 60 mL) that they should not be used with syringe infusion pumps. 3. POST a copy of this notification in your storeroom where the product is stored. 4. NOTIFY any customers to whom you may have distributed/forwarded affected product to or will send the product on to about this medical device product correction and share a copy of this notice. 5. RETURN the enclosed acknowledgment form via fax to 614-652-9648 or email to GMB [email protected], whether you have affected product or not. complete and return the provided acknowledgement form. For questions related to this notification and/or acknowledgement form that are not adequately addressed in this letter, please contact the market action team at: [email protected] or call 800-292-9332.
Quantity in commerce
3,113,846 units
Distribution
US Nationwide distribution to AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY.
Product codes and lots
Product Code: 1180100777; UDI/DI: 10192253034530 - each, 20192253034537 - box, 50192253034538 - case; Lot Numbers: 221201, 221202, 221203, 230201, 230202, 230203, 230204, 230205, 230601

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0145-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0145-2024Class IOngoingVoluntary: Firm initiated
Z-0146-2024Class IOngoingVoluntary: Firm initiated
Z-0147-2024Class IOngoingVoluntary: Firm initiated
Z-0148-2024Class IOngoingVoluntary: Firm initiated
Z-0149-2024Class IOngoingVoluntary: Firm initiated
Z-0150-2024Class IOngoingVoluntary: Firm initiated