Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0171-2024Class II

General Pack, REF BBGP31C

American Contract Systems, Inc. / initiated 2023-09-01 / Open, Classified / root cause: manufacturing process

American Contract Systems, Inc. began a recall of General Pack, REF BBGP31C on . FDA classified it as Class II and gives the reason as "During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0171-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
American Contract Systems, Inc.
Product
General Pack, REF BBGP31C
Product code
LRO General Surgery Tray
Reason for recall
During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 09/01/2023 by email. The notice explained the issue, risk, and requested the user discontinue the use of the I.V. Cath and Adhesive Dermabond and affix a warning label to each kit and case of affected product in inventory. The firm sent an updated letter on 09/22/2023 via email which corrected the product distribution date.
Quantity in commerce
57 cases
Distribution
US Nationwide distribution in the states of FL, IA.
Product codes and lots
UDI/DI 00191072166609, Lot Number 968231

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0171-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0171-2024Class IIOngoingVoluntary: Firm initiated
Z-0172-2024Class IIOngoingVoluntary: Firm initiated
Z-0173-2024Class IIOngoingVoluntary: Firm initiated
Z-0174-2024Class IIOngoingVoluntary: Firm initiated
Z-0175-2024Class IIOngoingVoluntary: Firm initiated
Z-0176-2024Class IIOngoingVoluntary: Firm initiated
Z-0177-2024Class IIOngoingVoluntary: Firm initiated
Z-0178-2024Class IIOngoingVoluntary: Firm initiated
Z-0179-2024Class IIOngoingVoluntary: Firm initiated
Z-0180-2024Class IIOngoingVoluntary: Firm initiated
Z-0181-2024Class IIOngoingVoluntary: Firm initiated
Z-0182-2024Class IIOngoingVoluntary: Firm initiated