Z-0127-2024Class II

TruDi Navigation System, Model: FG-2000-00, Catalog: ENS022B, when used with TruDi Curette, Models: TDC0005Z...

Acclarent Inc / initiated 2023-09-21 / Open, Classified / root cause: software

Acclarent, Inc. began a recall of TruDi Navigation System, Model: FG-2000-00, Catalog: ENS022B, when used with TruDi Curette, Models: TDC0005Z and TDC0005 on . FDA classified it as Class II and gives the reason as "When using affected curette and software, there is a discrepancy between the actual curette tip location and the location displayed on navigation systems intended for use during surgical procedures of the Ear Nose and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0127-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Acclarent Inc
Product
TruDi Navigation System, Model: FG-2000-00, Catalog: ENS022B, when used with TruDi Curette, Models: TDC0005Z and TDC0005
Product code
PGW Ear, Nose, And Throat Stereotaxic Instrument
Reason for recall
When using affected curette and software, there is a discrepancy between the actual curette tip location and the location displayed on navigation systems intended for use during surgical procedures of the Ear Nose and Throat (ENT) and ENT skull base surgery, which may cause delayed/prolonged surgery, cerebrospinal fluid leak, visual impairment, or skull base structural damage.
Root cause tag
software
FDA root cause
Software Design Change
Action
On 9/21/23, correction notices were mailed to customers who were asked to do the following: Please cease using the affected curette device with the affected navigation system until the new version of navigation system software version 2.3.2 is installed. 2. Please do not return any of the curettes within your inventory. 3. Complete and return the business response form via email to [email protected] 4. Forward this notice to any personnel in your facility that needs to be informed. 5. If any of the affected products have been forwarded to another facility, contact that facility and provide a copy of this to the relevant personnel. 6. Post a copy of this notice in a visible area for awareness If any adverse reactions or quality problems are experienced during the use of the affected navigation system with the curette, please report any issues to the firm via the product complaint phone number:1-877-775-2789, Option # 5.
Quantity in commerce
141
Distribution
US: CA, IL, WI, MA, TX, CO, FL, ID, KS, OH, OR, NE, MO, AZ, HI, TN, NY, ME, SC, NC, CT, DC, MN, LA, OK, PA, AL, IA, MI, SD
Product codes and lots
TruDi Navigation System, UDI-DI: 10846835018639, software version: V2.3.1 Update (2.3.1.144 and 2.3.1.166)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0127-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0127-2024Class IIOngoingVoluntary: Firm initiated