Z-0180-2024Class II

Sub Q Recorder Pack, REF BPRC87A

American Contract Systems, Inc. / initiated 2023-09-01 / Open, Classified / root cause: manufacturing process

American Contract Systems, Inc. began a recall of Sub Q Recorder Pack, REF BPRC87A on . FDA classified it as Class II and gives the reason as "During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0180-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
American Contract Systems, Inc.
Product
Sub Q Recorder Pack, REF BPRC87A
Product code
OES Cardiac Catheterization Kit
Reason for recall
During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 09/01/2023 by email. The notice explained the issue, risk, and requested the user discontinue the use of the I.V. Cath and Adhesive Dermabond and affix a warning label to each kit and case of affected product in inventory. The firm sent an updated letter on 09/22/2023 via email which corrected the product distribution date.
Quantity in commerce
96 cases
Distribution
US Nationwide distribution in the states of FL, IA.
Product codes and lots
UDI/DI 00191072187970, Lot Numbers: 969231

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0180-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0171-2024Class IIOngoingVoluntary: Firm initiated
Z-0172-2024Class IIOngoingVoluntary: Firm initiated
Z-0173-2024Class IIOngoingVoluntary: Firm initiated
Z-0174-2024Class IIOngoingVoluntary: Firm initiated
Z-0175-2024Class IIOngoingVoluntary: Firm initiated
Z-0176-2024Class IIOngoingVoluntary: Firm initiated
Z-0177-2024Class IIOngoingVoluntary: Firm initiated
Z-0178-2024Class IIOngoingVoluntary: Firm initiated
Z-0179-2024Class IIOngoingVoluntary: Firm initiated
Z-0180-2024Class IIOngoingVoluntary: Firm initiated
Z-0181-2024Class IIOngoingVoluntary: Firm initiated
Z-0182-2024Class IIOngoingVoluntary: Firm initiated