Cath Cardiac Pack, REF IHCC03V
American Contract Systems, Inc. began a recall of Cath Cardiac Pack, REF IHCC03V on . FDA classified it as Class II and gives the reason as "During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0182-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- American Contract Systems, Inc.
- Product
- Cath Cardiac Pack, REF IHCC03V
- Product code
- NWU Endoscope Introducer Kit
- Reason for recall
- During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 09/01/2023 by email. The notice explained the issue, risk, and requested the user discontinue the use of the I.V. Cath and Adhesive Dermabond and affix a warning label to each kit and case of affected product in inventory. The firm sent an updated letter on 09/22/2023 via email which corrected the product distribution date.
- Quantity in commerce
- 1260 units
- Distribution
- US Nationwide distribution in the states of FL, IA.
- Product codes and lots
- UDI/DI 00191072189431, Lot Numbers: 724221, 679221, 945231
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0171-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0172-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0173-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0174-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0175-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0176-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0177-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0178-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0179-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0180-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0181-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0182-2024 | Class II | Ongoing | Voluntary: Firm initiated |