Portex Pro-Vent Arterial Blood Sampling Kit with Dry Lithium Heparin for Gases and Electrolytes, REF 4599P-1
Smiths Medical ASD Inc. began a recall of Portex Pro-Vent Arterial Blood Sampling Kit with Dry Lithium Heparin for Gases and Electrolytes, REF 4599P-1 on . FDA classified it as Class II and gives the reason as "The complaint involved the absence of a component called "ASM Filter-Pro ABG SYR CAP" from the packaging of Pro-Vent Arterial Blood Sampling (ABG) Kits. Specifically, this issue affected a total of 31,900 kits under the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0141-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Medical
- Product
- Portex Pro-Vent Arterial Blood Sampling Kit with Dry Lithium Heparin for Gases and Electrolytes, REF 4599P-1
- Product code
- JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
- Reason for recall
- The complaint involved the absence of a component called "ASM Filter-Pro ABG SYR CAP" from the packaging of Pro-Vent Arterial Blood Sampling (ABG) Kits. Specifically, this issue affected a total of 31,900 kits under the FGSKU# 4599P-1 and Lot# 4331283. Tubes without caps have the potential to generate delayed or erroneous test results and/or exposure to infectious blood.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Smiths Medical issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 09/06/2023 by mail. The notice explained the problem with the device, potential risk, and requested the following: 1. Discontinue use, distribution, and quarantine all affected product in inventory. 2. Notify all potential users within the consignee organization 3. Distributors were directed to notify their customers. Questions: Customer Service 1-(800)-258-5361
- Quantity in commerce
- 31900 devices
- Distribution
- US
- Product codes and lots
- UDI/DI 30351688028635, Lot Number 4331283
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0141-2024 | Class II | Ongoing | Voluntary: Firm initiated |