EMPOWR Acetabular System, Liner, 10 Hooded, HXe+, 40H, REF: 942-01-40H
Encore Medical, LP began a recall of EMPOWR Acetabular System, Liner, 10 Hooded, HXe+, 40H, REF: 942-01-40H on . FDA classified it as Class II and gives the reason as "Knee insert and acetabular system parts were swapped within inner pack prior to the product being sealed within the tray, which could result in surgeons not having the correct product for the surgery due to different...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0103-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Encore Medical, LP
- Product
- EMPOWR Acetabular System, Liner, 10 Hooded, HXe+, 40H, REF: 942-01-40H
- Product code
- LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
- Reason for recall
- Knee insert and acetabular system parts were swapped within inner pack prior to the product being sealed within the tray, which could result in surgeons not having the correct product for the surgery due to different anatomies (Hip vs Knee).
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- On 8/22/23, recall notices were emailed to agents who were asked to do the following: 1) Complete acknowledgement and response form and contact Customer Service at 1-800-456-8696 to place a replacement order and receive an RMA number. 2) The notice needs to be passed on to all those within your organization or to any organization where the potentially affected product has been transferred. If you have any questions, email [email protected]
- Quantity in commerce
- 20
- Distribution
- US Nationwide distribution in the states of FL, CA, IN, HI, LA, MN, AZ, TX, NJ, PA, KS, TN, PR, SC, WA.
- Product codes and lots
- UDI-DI: 00190446309833, Lot: 629Z1114, Expiration: March 30, 2028
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0102-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0103-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.