Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0103-2024Class II

EMPOWR Acetabular System, Liner, 10 Hooded, HXe+, 40H, REF: 942-01-40H

Encore Medical, LP / initiated 2023-08-22 / Open, Classified / root cause: labeling and UDI

Encore Medical, LP began a recall of EMPOWR Acetabular System, Liner, 10 Hooded, HXe+, 40H, REF: 942-01-40H on . FDA classified it as Class II and gives the reason as "Knee insert and acetabular system parts were swapped within inner pack prior to the product being sealed within the tray, which could result in surgeons not having the correct product for the surgery due to different...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0103-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Encore Medical, LP
Product
EMPOWR Acetabular System, Liner, 10 Hooded, HXe+, 40H, REF: 942-01-40H
Product code
LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented
Reason for recall
Knee insert and acetabular system parts were swapped within inner pack prior to the product being sealed within the tray, which could result in surgeons not having the correct product for the surgery due to different anatomies (Hip vs Knee).
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
On 8/22/23, recall notices were emailed to agents who were asked to do the following: 1) Complete acknowledgement and response form and contact Customer Service at 1-800-456-8696 to place a replacement order and receive an RMA number. 2) The notice needs to be passed on to all those within your organization or to any organization where the potentially affected product has been transferred. If you have any questions, email [email protected]
Quantity in commerce
20
Distribution
US Nationwide distribution in the states of FL, CA, IN, HI, LA, MN, AZ, TX, NJ, PA, KS, TN, PR, SC, WA.
Product codes and lots
UDI-DI: 00190446309833, Lot: 629Z1114, Expiration: March 30, 2028