Z-0184-2024Class I

BD Alaris PCU REF 8015

CareFusion 303, Inc / initiated 2023-09-15 / Open, Classified / root cause: labeling and UDI

CareFusion 303, Inc. began a recall of BD Alaris PCU REF 8015 on . FDA classified it as Class I and gives the reason as "Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0184-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
CareFusion 303, Inc
Product
BD Alaris PCU REF 8015
Product code
FRN Pump, Infusion
Reason for recall
Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
On September 15, 2023, CareFusion a subsidiary of Becton Dickinson (BD) posted an "Urgent Medical Device Labeling Correction" on their website. On October 4, 2023 CareFusion mailed a copy of the correction notification via FedEx. CareFusion asked consignees to take the following actions: 1. You should cease use of Cardinal Health branded Monoject" syringes with the BD Alaris" Syringe and PCA Modules; except for SKU 8881135609, which is not impacted by this notification. 2. Please refer to Attachments 1 and 2 of this notice for a full selection of syringes that are compatible with the BD Alaris" Syringe and PCA Modules. 3. Replace your current BD Alaris" Syringe and PCA Module Compatibility Lists with Attachments 1 and 2 of this notice. 4. Please circulate this notice to all those who require awareness within your organization or to any organization whom you may have transferred BD Alaris" Infusion System devices to. 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification. 6. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787
Quantity in commerce
867,362 units
Distribution
Worldwide Distribution: US (nationwide): AK, AL, AR, AS, AZ, CA, CO, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MP, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY; OUS (foreign): Canada, Germany, Cuba, Italy, Japan, South Korea, Turkey, United Arab Emirates, Australia, Bahrain, Botswana, Kuwait, Mexico, Netherlands, Philippines, Pakistan, Qatar, South Africa, Singapore, Belgium, Israel, India, New Zealand, Malaysia, Taiwan, Switzerland
Product codes and lots
All Lots/ UDI DI: 10885403516030,10885403515316,10885403812033,10885403515286,10885403812026, 10885403515293,10885403494291,10885403515309,10885403812002

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0184-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0184-2024Class IOngoingVoluntary: Firm initiated
Z-0185-2024Class IOngoingVoluntary: Firm initiated
Z-0186-2024Class IOngoingVoluntary: Firm initiated