Z-0206-2024Class II

RayCare 5B and 6A including service packs- An oncology information system used to support workflows and...

RaySearch Laboratories AB (publ) / initiated 2023-09-12 / Open, Classified

RAYSEARCH LABORATORIES AB began a recall of RayCare 5B and 6A including service packs- An oncology information system used to support workflows and scheduling, as well as clinical information, planning... on . FDA classified it as Class II and gives the reason as "An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible for documents in a read-only state when opened outside the Documents workspace when using document fields configured with...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0206-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
RaySearch Laboratories AB (publ)
Product
RayCare 5B and 6A including service packs- An oncology information system used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care Product name (build number): (1)RayCare 5B SP1 (5.1.1.60246); (2)RayCare 5B SP3 (5.1.3.60023); (3) RayCare 6A (6.0.0.60553)
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible for documents in a read-only state when opened outside the Documents workspace when using document fields configured with single- or multi-selectable values.
FDA root cause
Under Investigation by firm
Action
Raysearch issued Field Safety Notice, Medical Device Correction #121600 via email on 9/12/23. Letter states reason for recall, health risk and action to take: Educate staff and all users to open configurable documents from the Documents workspace to avoid the issue. " Review all configured documents and consider reconfiguration in a way that will not trigger the issue. " Inspect your product and identify all installed units with the above software version number(s). " Confirm you have read and understood this notice by replying to the notification email. Solution This issue will be resolved in the next version of RayCare, scheduled for market release in December 2023 (RayCare 2024A) (subject to market clearance in some markets). If customers wish to continue using versions of RayCare affected by this notice, all users must maintain awareness of this notice. Alternatively, customers can choose to upgrade to the new version once it becomes available for clinical use. For regulatory information, please contact [email protected].
Quantity in commerce
2 sytems
Distribution
MD, TN
Product codes and lots
UDI-DI: (1)RayCare 5B SP1 (5.1.1.60246) 0735000201053220220316; (2)RayCare 5B SP3 (5.1.3.60023) 0735000201069320221027; (3) RayCare 6A (6.0.0.60553) 0735000201056320220617

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0206-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0206-2024Class IIOngoingVoluntary: Firm initiated