LEAD 3387S-40 STIMLOC DBS, Model Number 3387S-40. Electrical implantable device
Medtronic Neuromodulation began a recall of LEAD 3387S-40 STIMLOC DBS, Model Number 3387S-40. Electrical implantable device on . FDA classified it as Class II and gives the reason as "There are a small number of units of the DBS Lead with StimLoc kit (3387S-40) due to a discrepancy in the Use By Date (UBD) printed on the outer packaged kit label versus the UBD printed on the individual product label". As of , the recall status on record is Completed, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0217-2024
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Medtronic Neuromodulation
- Product
- LEAD 3387S-40 STIMLOC DBS, Model Number 3387S-40. Electrical implantable device.
- Product code
- MHY Stimulator, Electrical, Implanted, For Parkinsonian Tremor
- Reason for recall
- There are a small number of units of the DBS Lead with StimLoc kit (3387S-40) due to a discrepancy in the Use By Date (UBD) printed on the outer packaged kit label versus the UBD printed on the individual product label.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- Medtronic issued a Medical Device Removal notice to its consignees on 05/27/2023 by US mail. The notice explained the issue, potential risk, and requested the return of the affected product.
- Quantity in commerce
- 14 units
- Distribution
- US: MO, FL, TX, WA
- Product codes and lots
- UDI/DI 00763000274320, Serial Number VA2JV2V
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0217-2024 | Class II | Completed | Voluntary: Firm initiated |