Z-0151-2024Class I

Novum IQ Syringe infusion system, Product Code 40800BAXUS

Baxter Healthcare Corporation / initiated 2023-10-13 / Open, Classified

Baxter Healthcare Corporation began a recall of Novum IQ Syringe infusion system, Product Code 40800BAXUS on . FDA classified it as Class I and gives the reason as "Baxter is issuing an Urgent Medical Device Correction for the Novum IQ Syringe Pump. Baxter identified that after multiple downstream occlusion alarms, the pump may display an Infusion Complete alarm even though...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0151-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Baxter Healthcare Corporation
Product
Novum IQ Syringe infusion system, Product Code 40800BAXUS
Product code
FRN Pump, Infusion
Reason for recall
Baxter is issuing an Urgent Medical Device Correction for the Novum IQ Syringe Pump. Baxter identified that after multiple downstream occlusion alarms, the pump may display an Infusion Complete alarm even though uninfused fluid remains in the syringe.
FDA root cause
Under Investigation by firm
Action
Baxter issued an "Urgent Medical Device Correction" notice on 10/13/2023 via USPS first-class mail. The notice explained the issue, hazard involved, and provided mitigation procedures in accordance with the Operator's Manual until a software upgrade becomes available. For additional questions or experience quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-843-7867, Monday through Friday, between 7:00 am and 7:00 pm Eastern Time.
Quantity in commerce
2023 units
Distribution
US Nationwide Distribution to states of: AR, AZ, CA, FL, GA, HI, IA, IN, KS, LA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, PA, TX, UT, VA, VT, WA, WI AND WV.
Product codes and lots
All Serial Numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0151-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0151-2024Class IOngoingVoluntary: Firm initiated