Monitron II Waveform Analyzer, REF: S00015-1, used with ventilators
Percussionaire Corporation began a recall of Monitron II Waveform Analyzer, REF: S00015-1, used with ventilators on . FDA classified it as Class II and gives the reason as "Note this recall occurred in 2020 and 2021. Reports of screen freeze on visual displays of waveform analyzers that are paired with ventilators. Subsequently, waveform analyzers were serviced to install PAL...". As of , the recall status on record is Completed, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0259-2024
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Percussionaire Corp.
- Product
- Monitron II Waveform Analyzer, REF: S00015-1, used with ventilators.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Note this recall occurred in 2020 and 2021. Reports of screen freeze on visual displays of waveform analyzers that are paired with ventilators. Subsequently, waveform analyzers were serviced to install PAL microprocessor chips and firmware.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Waveform analyzers were serviced from September 2020 through September 2021.
- Quantity in commerce
- 212
- Distribution
- US: CA, ND, MS, MD
- Product codes and lots
- UDI-DI: 00849436000068, Serial Numbers: WF21150, WF21159, WF21163, WF21164, WF21167, WF21171, WF21371, WF21372, WF21377, WF21378, WF21158, WF21391, WF21165, WF21166, WF21168, WF21169, WF21170, WF21172, WF21173, WF21174, WF21175, WF21176, WF21177, WF21178, WF21179, WF21182, WF21183, WF21184, WF21185, WF21186, WF21187, WF21188, WF21189, WF21190, WF21192, WF21193, WF21194, WF21195, WF21196, WF21197, WF21198, WF21199, WF21200, WF21201, WF21202, WF21203, WF21204, WF21205, WF21206, WF21207, WF21208, WF21209, WF21210, WF21211, WF21212, WF21213, WF21214, WF21215, WF21217, WF21218, WF21219, WF21220, WF21221, WF21222, WF21223, WF21224, WF21225, WF21226, WF21227, WF21228, WF21229, WF21230, WF21231, WF21232, WF21233, WF21234, WF21235, WF21236, WF21237, WF21238, WF21239, WF21240, WF21241, WF21242, WF21243, WF21244, WF21245, WF21246, WF21247, WF21248, WF21249, WF21250, WF21251, WF21252, WF21253, WF21254, WF21255, WF21256, WF21257, WF21258, WF21259, WF21260, WF21261, WF21262, WF21263, WF21264, WF21265, WF21266, WF21267, WF21268, WF21269, WF21270, WF21271, WF21272, WF21273, WF21274, WF21275, WF21276, WF21277, WF21278, WF21279, WF21280, WF21281, WF21282, WF21283, WF21284, WF21285, WF21286, WF21287, WF21288, WF21289, WF21290, WF21291, WF21292, WF21293, WF21294, WF21295, WF21296, WF21297, WF21298, WF21299, WF21300, WF21301, WF21302, WF21303, WF21304, WF21305, WF21306, WF21307, WF21308, WF21309, WF21310, WF21311, WF21312, WF21313, WF21314, WF21315, WF21316, WF21317, WF21318, WF21319, WF21320, WF21321, WF21322, WF21323, WF21324, WF21325, WF21326, WF21327, WF21328, WF21329, WF21330, WF21331, WF21332, WF21333, WF21334, WF21335, WF21336, WF21337, WF21338, WF21339, WF21340, WF21341, WF21342, WF21343, WF21344, WF21345, WF21346, WF21347, WF21348, WF21349, WF21350, WF21351, WF21352, WF21353, WF21354, WF21355, WF21356, WF21357, WF21358, WF21359, WF21360, WF21361, WF21362, WF21363, WF21364, WF21365, WF21366, WF21367, WF21368, WF21369, WF21370
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0259-2024 | Class II | Completed | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.