Z-0258-2024Class II
DJO EMPOWR Knee Punch Handle, REF: 801-05-040
Encore Medical, LP / initiated 2023-09-29 / Open, Classified
Encore Medical, LP began a recall of DJO EMPOWR Knee Punch Handle, REF: 801-05-040 on . FDA classified it as Class II and gives the reason as "Knee punch handle tip is breaking during surgery due to the lack of heat treatment, which may lead to a disruption in surgery". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0258-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Encore Medical, LP
- Product
- DJO EMPOWR Knee Punch Handle, REF: 801-05-040
- Product code
- MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
- Reason for recall
- Knee punch handle tip is breaking during surgery due to the lack of heat treatment, which may lead to a disruption in surgery.
- FDA root cause
- Process change control
- Action
- On 9/29/23, recall notices were emailed to customers who were asked to do the following: 1) Knee punch handles should be visually inspected prior to use for any damage. Upon completion of surgery, visually inspect and verify there is no breakage. The attached quality bulletin provides guidance on an acceptable and unacceptable device. 2) Contact your customer service rep at [email protected] to report failures or to request replacements. 3) Complete and return the response form that asks customers to check all related devices in inventory and report the number of related devices in inventory. Questions can be directed to [email protected]
- Quantity in commerce
- 406
- Distribution
- US: VA, TN, MN, TX, IL, IN, KY, OH, SC, MS, RI, ME, NJ, NY, MD, CA, KS, LA, PR, AL, FL, GA, AZ, WA
- Product codes and lots
- UDI-DI: 00190446238881, Lots: 305197L02D, 305197L02E, 318871L10D, 318871L10E, 305197L02B, 305197L02C, 318871L10B, 318871L10C, 325045L17A, 325045L17B, 325045L17C, 305197L02, 318871L10, 325045L17, 327669L03, 331146L01, 332158L06, 334314L04, 336735L04, 336801L01, 340391L01, 341961L01, 342586L01, 346101L01, 350076L01, 351106L02, 360277L01, 366833L03, 374648L01, 380510L01
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0258-2024 | Class II | Ongoing | Voluntary: Firm initiated |