Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0258-2024Class II

DJO EMPOWR Knee Punch Handle, REF: 801-05-040

Encore Medical, LP / initiated 2023-09-29 / Open, Classified

Encore Medical, LP began a recall of DJO EMPOWR Knee Punch Handle, REF: 801-05-040 on . FDA classified it as Class II and gives the reason as "Knee punch handle tip is breaking during surgery due to the lack of heat treatment, which may lead to a disruption in surgery". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0258-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Encore Medical, LP
Product
DJO EMPOWR Knee Punch Handle, REF: 801-05-040
Product code
MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer
Reason for recall
Knee punch handle tip is breaking during surgery due to the lack of heat treatment, which may lead to a disruption in surgery.
FDA root cause
Process change control
Action
On 9/29/23, recall notices were emailed to customers who were asked to do the following: 1) Knee punch handles should be visually inspected prior to use for any damage. Upon completion of surgery, visually inspect and verify there is no breakage. The attached quality bulletin provides guidance on an acceptable and unacceptable device. 2) Contact your customer service rep at [email protected] to report failures or to request replacements. 3) Complete and return the response form that asks customers to check all related devices in inventory and report the number of related devices in inventory. Questions can be directed to [email protected]
Quantity in commerce
406
Distribution
US: VA, TN, MN, TX, IL, IN, KY, OH, SC, MS, RI, ME, NJ, NY, MD, CA, KS, LA, PR, AL, FL, GA, AZ, WA
Product codes and lots
UDI-DI: 00190446238881, Lots: 305197L02D, 305197L02E, 318871L10D, 318871L10E, 305197L02B, 305197L02C, 318871L10B, 318871L10C, 325045L17A, 325045L17B, 325045L17C, 305197L02, 318871L10, 325045L17, 327669L03, 331146L01, 332158L06, 334314L04, 336735L04, 336801L01, 340391L01, 341961L01, 342586L01, 346101L01, 350076L01, 351106L02, 360277L01, 366833L03, 374648L01, 380510L01

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0258-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0258-2024Class IIOngoingVoluntary: Firm initiated