RAPIDPOINT 500e Blood Gas System

Siemens Healthcare Diagnostics Inc. / initiated 2023-09-26 / Open, Classified

Siemens Healthcare Diagnostics Inc began a recall of RAPIDPOINT 500e Blood Gas System on . FDA classified it as Class II and gives the reason as "There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0422-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
RAPIDPOINT 500e Blood Gas System
Product code
JGS Electrode, Ion Specific, Sodium
Reason for recall
There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The erroneous sodium levels with positive bias may lead to unrecognized hyponatremia and/or the inappropriate treatment of hypernatremia may lead to iatrogenic hyponatremia, resulted in a delay in patient diagnosis and optimal patient management.
FDA root cause
Under Investigation by firm
Action
The firm issued an Urgent Medical Device Correction notice on 9/26/2023 to customers asking that customers review the notice with the facility's Medical Director to determine the appropriate course of action for previously generated results, complete and return the Field Correction Effectiveness Check Form within 30 days, forward this notice to customers if your firm is a distributor, and retain the notification in your system's Operator's Guide and forward to those who may have received product. The Field Correction Effectiveness Check Form is to be returned via email to [email protected] or by fax to 312-275-7795. Questions are to be directed to your local Siemens Healthineers Technical Support Representative.
Quantity in commerce
8,749 units
Distribution
Domestic: Nationwide Distribution. Foreign: Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Congo, Republic, Costa Rica, Cote dIvoire, Croatia, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Fiji, Finland, France, French, Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Lesotho, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Martinique, Mexico, Moldova, Montenegro, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestinian Territory, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom Great Britain, Uruguay, & Vietnam.
Product codes and lots
Material No. 11416751, 11416752, 11416754, & 11416755; UDI-DI: 00630414286150, 00630414286167, 00630414286174, & 00630414286143; All Units and Software Versions.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0422-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0422-2024Class IIOngoingVoluntary: Firm initiated
Z-0423-2024Class IIOngoingVoluntary: Firm initiated