VariSoft infusion set, Model Number 1002827
Unomedical A/S began a recall of VariSoft infusion set, Model Number 1002827 on . FDA classified it as Class I and gives the reason as "The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0232-2024
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Unomedical A/S
- Product
- VariSoft infusion set, Model Number 1002827
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.
- Root cause tag
- manufacturing process
- FDA root cause
- Process change control
- Action
- Unomedical a/s issued an Urgent Field Safety Notice to its sole consignee on 10/11/2023 via email. The notice explained the issue, potential risk to patient, and requested the distributor to quarantine all affected product pending return and to notify their customers.
- Quantity in commerce
- 8028 units
- Distribution
- US Nationwide distribution in the state of CA.
- Product codes and lots
- GTIN 05705244018709, Serial Numbers: 5388357, 5388362, 5388372, 5388366; GTIN 05705244018693, Serial Numbers: 5388357, 5388362, 5388372, 5388366
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0232-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0233-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0234-2024 | Class I | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
