VariSoft infusion set, Model Number 1002830

Unomedical A/S / initiated 2023-10-11 / Open, Classified / root cause: manufacturing process

Unomedical A/S began a recall of VariSoft infusion set, Model Number 1002830 on . FDA classified it as Class I and gives the reason as "The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0234-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Unomedical A/S
Product
VariSoft infusion set, Model Number 1002830
Product code
FPA Set, Administration, Intravascular
Reason for recall
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.
Root cause tag
manufacturing process
FDA root cause
Process change control
Action
Unomedical a/s issued an Urgent Field Safety Notice to its sole consignee on 10/11/2023 via email. The notice explained the issue, potential risk to patient, and requested the distributor to quarantine all affected product pending return and to notify their customers.
Quantity in commerce
3362 units
Distribution
US Nationwide distribution in the state of CA.
Product codes and lots
GTIN 05705244018792, Serial Numbers: 5388368; GTIN 05705244018785, Serial Numbers: 5388368

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0234-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0232-2024Class IOngoingVoluntary: Firm initiated
Z-0233-2024Class IOngoingVoluntary: Firm initiated
Z-0234-2024Class IOngoingVoluntary: Firm initiated