Mobilett Mira wireless (VA20) mobile x-ray system(s)
Siemens Medical Solutions USA, Inc began a recall of Mobilett Mira wireless (VA20) mobile x-ray system(s) on . FDA classified it as Class II and gives the reason as "Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0223-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Mobilett Mira wireless (VA20) mobile x-ray system(s)
- Product code
- IZL System, X-Ray, Mobile
- Reason for recall
- Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual indicator (exposure LED light) the operator can monitor the x ray release. Also, after a successful x ray exposure, the clinical image is displayed on the monitor.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Customer notification by means of a Customer Safety Advisory Notice (via electronic mail or FedEx) alerting customers to the lack of acoustic signal issue and providing instructions to monitor the visual indicator (exposure LED light) to avoid additional, unnecessary x-ray dose. The CSAN also includes the plan for the affected D810 boards. The boards (containing the faulty mounting of the buzzer) will be replaced, free of charge, at the customer site with field corrective action XP042/23/P. There is no software update.
- Quantity in commerce
- 4 distributed in U.S.
- Distribution
- US Nationwide - Worldwide
- Product codes and lots
- Model Number 10273100 / UDI 04056869009117 Serial 1511 2168 2525 2633
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0223-2024 | Class II | Ongoing | FDA Mandated |