Z-0223-2024Class II

Mobilett Mira wireless (VA20) mobile x-ray system(s)

Siemens Medical Solutions USA, Inc. / initiated 2023-10-05 / Open, Classified / root cause: manufacturing process

Siemens Medical Solutions USA, Inc began a recall of Mobilett Mira wireless (VA20) mobile x-ray system(s) on . FDA classified it as Class II and gives the reason as "Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0223-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Mobilett Mira wireless (VA20) mobile x-ray system(s)
Product code
IZL System, X-Ray, Mobile
Reason for recall
Due to the incorrect mounting of the buzzer component, there is no acoustic signal during x ray activation. Otherwise, the system is fully functional, and x ray is applied correctly. By means of the additional visual indicator (exposure LED light) the operator can monitor the x ray release. Also, after a successful x ray exposure, the clinical image is displayed on the monitor.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Customer notification by means of a Customer Safety Advisory Notice (via electronic mail or FedEx) alerting customers to the lack of acoustic signal issue and providing instructions to monitor the visual indicator (exposure LED light) to avoid additional, unnecessary x-ray dose. The CSAN also includes the plan for the affected D810 boards. The boards (containing the faulty mounting of the buzzer) will be replaced, free of charge, at the customer site with field corrective action XP042/23/P. There is no software update.
Quantity in commerce
4 distributed in U.S.
Distribution
US Nationwide - Worldwide
Product codes and lots
Model Number 10273100 / UDI 04056869009117 Serial 1511 2168 2525 2633

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0223-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0223-2024Class IIOngoingFDA Mandated