Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0408-2024Class II

Lugol s Iodine Solution 500ML, Part Number 624-71

EMD Millipore Corporation / initiated 2023-10-17 / Open, Classified / root cause: packaging

EMD Millipore Corporation began a recall of Lugol s Iodine Solution 500ML, Part Number 624-71 on . FDA classified it as Class II and gives the reason as "Due to the incorrect packaging utilized with lot 3124, there' s an improper seal between the cap and bottle. With an improper seal the Assay (current specification of 24.8 to 27.2 mL) is expected to continue to drop...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0408-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
EMD Millipore Corporation
Product
Lugol s Iodine Solution 500ML, Part Number 624-71
Product code
IAM Solution, Lugol'S
Reason for recall
Due to the incorrect packaging utilized with lot 3124, there' s an improper seal between the cap and bottle. With an improper seal the Assay (current specification of 24.8 to 27.2 mL) is expected to continue to drop over time as iodine vapors continue to dissipate from the bottles.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
A Product Recall notification letter dated 10/17/23 was sent to customers. Required/Recommended Action Please immediately examine your inventory to determine if you have any remaining impacted product and forward a copy of this letter to the personnel in your organization that utilize this product and/or need to be made aware of this issue. Please discard this material and request a replacement. Please complete attachment 1 form acknowledging receipt of this letter and return to the noted email address. In addition, if you may have further distributed this product, please identify your customers and notify them at once of this issue. Your notification to customers should include a copy of this notification. You may contact Technical or Customer Service to discuss an appropriate solution for your needs. If you would like more information about this FCA notification, please contact your local representative or use the link www.sigmaaldrich.com/offices for our website to obtain the phone number and/or email address of your local office.
Quantity in commerce
5 units
Distribution
US States: CA, NJ, WV
Product codes and lots
Lot Number 3124

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0408-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0408-2024Class IIOngoingVoluntary: Firm initiated