Z-0448-2024Class II
Monarch Platform REF MON-000005-01, 100-240V- 50/60Hz 1440w
Auris Health, Inc / initiated 2023-10-18 / Open, Classified / root cause: software
AURIS HEALTH INC began a recall of Monarch Platform REF MON-000005-01, 100-240V- 50/60Hz 1440w on . FDA classified it as Class II and gives the reason as "Their is the potential that software issues may result in flipped image of the virtual Bronchoscope view". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0448-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Auris Health, Inc
- Product
- Monarch Platform REF MON-000005-01, 100-240V- 50/60Hz 1440w
- Product code
- EOQ Bronchoscope (Flexible Or Rigid)
- Reason for recall
- Their is the potential that software issues may result in flipped image of the virtual Bronchoscope view.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- On October 19, 2023, Auris Health, Inc. issued an "URGENT: MEDICAL DEVICE CORRECTION" notification to affected consignees. Auris asked consignees to take the following actions: Should an image flip be detected in the Virtual Bronchoscope view while navigating toward the target, please perform the following steps to correct the image orientation: 1. Retract the MONARCH Bronchoscope outside the patient 2. Undock the MONARCH Bronchoscope from the arms 3. Clear any fault pop-up messages 4. Re-dock the MONARCH Bronchoscope to the arms 5. Re-initialize Navigation and proceed with the case 6. Please review this letter carefully and share it with anyone in your facility that needs to be informed. 7. Please complete all fields of the attached Acknowledgement Form confirming receipt of this notification and return the form via email to [email protected]. If you have any questions regarding this letter or need to report any customer complaints, please contact the MONARCH Support team at +1 (800) 434-0032 or via email at [email protected].
- Quantity in commerce
- 110 units In total
- Distribution
- US Distribution: AK, AL, AZ, CA, CO, CT, DC, DE, FL, GA, IL, IN, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OR, PA, SC, SD, TN, TX, WI & WV.
- Product codes and lots
- Software Version 2.2.4
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0448-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0449-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0450-2024 | Class II | Ongoing | Voluntary: Firm initiated |