Z-0563-2024Class I

(1) Nurse Kit, SKU #EDUC1025, Case UPC 40195327196647, Non-sterile, For Training Purposes Only, Not For Human...

MEDLINE INDUSTRIES, LP - Northfield / initiated 2023-10-16 / Open, Classified / root cause: packaging

MEDLINE INDUSTRIES, LP - Northfield began a recall of (1) Nurse Kit, SKU #EDUC1025, Case UPC 40195327196647, Non-sterile, For Training Purposes Only, Not For Human Use.; and (2) Nursing Tote Kit, SKU #EDUC5056A... on . FDA classified it as Class I and gives the reason as "The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-0563-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
(1) Nurse Kit, SKU #EDUC1025, Case UPC 40195327196647, Non-sterile, For Training Purposes Only, Not For Human Use.; and (2) Nursing Tote Kit, SKU #EDUC5056A, Case UPC 40195327153695, Non-sterile, For Training Purposes Only, Not For Human Use.
Product code
NSB Patient Personal Hygiene Kit
Reason for recall
The component tracheostomy brush has a sharp edge at the tip that can potentially puncture tubing during use and/or cause user injury. Additionally, there have been incidents of the tracheostomy brush bristles detaching prior to use in the packaging and/or during use.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
The recalling firm issued "MEDICAL DEVICE RECALL" letters dated 10/16/2023 via first class mail and email explaining the reason for recall and the required actions that needed to be taken. The actions for consignees to take included checking their stock for the affected item numbers and affected lot numbers which are located within the recall portal; using the link, https://recalls.medline.com, provided in the letter to complete the response form with the quantity of affected product in inventory; and destroy the affected inventory for credit. If the consignee is a distributor who has resold or transferred the product to another company or individual, they are to notify them of this recall and document the amount of product their customer destroyed on their (the distributor) response form. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Quantity in commerce
550 kits
Distribution
US Nationwide distribution including Puerto Rico. OUS (Foreign) distribution was made to Canada, Panama, and United Arab Emirates.
Product codes and lots
(1) EDUC1025 - Lot numbers 22JDB315 and 22HDC340; and (2) EDUC0506A - Lot numbers 22HDC201 and 22HDA420.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0563-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0561-2024Class IOngoingVoluntary: Firm initiated
Z-0562-2024Class IOngoingVoluntary: Firm initiated
Z-0563-2024Class IOngoingVoluntary: Firm initiated