with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to...
Breas Medical, Inc. began a recall of with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to ventilator device, Vivo 45LS. intended for partial... on . FDA classified it as Class II and gives the reason as "Produced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing patients using MPV-VCV and MPV-PCV modes of ventilation". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0405-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Breas Medical AB
- Product
- with Notch 22M/15F sold as part of Breas Medical MPV circuit 2 pc Kits: 006093, 006653, an accessory to ventilator device, Vivo 45LS. intended for partial insertion into a patient's mouth (in between the lips) to facilitate access to the respiratory system.
- Product code
- NOU Continuous, Ventilator, Home Use
- Reason for recall
- Produced with unintended open slits on the side of the mouthpiece body, may result in insufficient ventilation in spontaneous breathing patients using MPV-VCV and MPV-PCV modes of ventilation
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Breas Medical contacted customers via telephone followed by email on 10/13/23. Letter (Breas reference: SCAR-330) states reason for recall, health risk and action to take: Breas Medical recommends that all those potentially affected by this notice inspect existing inventory and identify mouthpieces which have been set up with the Vivo 45 LS for patient use. All defective mouthpieces shall be discarded. Breas Medical will provide replacement NIV Angled Mouthpieces, article number 008234, for all defective inventory and for devices in patient possession, where applicable. To obtain replacement NIV Angled Mouthpieces, we are asking customers to identify mouthpieces delivered from Breas since April 4, 2023 and complete the response form. We kindly ask you to acknowledge receipt of this Field Safety Notice together with your choice to remedy the issue using attached reply form NIV Angled Mouthpiece Field Safety Notice Distributor/Customer Reply Form. Contact Information Email: [email protected] Helpline: +1 (855) 436-8724 Postal Address 16 Esquire Road, N. Billerica, MA 01862 Web Portal: breas.com
- Quantity in commerce
- 215 units
- Distribution
- Nationwide Foreign: Canada
- Product codes and lots
- UDI-DI: 05030267120505 NIV Angled Mouthpiece with Notch 22M/15F; 07321820060931 MPV Kit 006093; 07321820066537 MPV Kit 006653 MPV mouthpiece Lot Numbers: 32304444, 32305269 and 32306017
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0405-2024 | Class II | Ongoing | Voluntary: Firm initiated |