Z-0523-2024Class II

ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100

Siemens Medical Solutions USA, Inc. / initiated 2023-10-17 / Open, Classified / root cause: design

Siemens Medical Solutions USA, Inc began a recall of ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100 on . FDA classified it as Class II and gives the reason as "Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0523-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100
Product code
OWB Interventional Fluoroscopic X-Ray System
Reason for recall
Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and switched on. During this time, system movement is only possible with the "override" function with very slow speed and without further collision protection. This may lead to temporary unavailability of the system for diagnostic and therapeutic usage.
Root cause tag
design
FDA root cause
Component design/selection
Action
On October 17, 2023, the firm issued an URGENT MEDICAL DEVICE CORRECTION letter to affected customers. According to the firm's letter, as the unintended behavior is caused by thermal startup effects, its occurrence can be avoided by only switching off the system for a short time. The firm suggests keeping the system running (except during proposed weekly reboot) if availability for treatment of emergency patients is necessary. Please plan for up to half an hour warm-up time after switching on the system. The firm will contact affected customers to modify the mounting of the tube cover to mitigate the probability of occurrence of the issue. You may contact the firm's service organization for an earlier appointment at 1-800-888-7436.
Quantity in commerce
38 worldwide, 11 US
Distribution
US Nationwide - Worldwide Distribution: Domestic distribution to FL GA IA IL IN MI MN MO MT PA TX. Foreign distribution to Australia Canada Switzerland China Germany Denmark Egypt Spain Finland United Kingdom Israel Netherlands Sweden South Africa
Product codes and lots
UDI-DI 04056869295923 Serial Numbers: All SN's associated with material number 11328100

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0523-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0523-2024Class IIOngoingVoluntary: Firm initiated