ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100
Siemens Medical Solutions USA, Inc began a recall of ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100 on . FDA classified it as Class II and gives the reason as "Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0523-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- ARTIS icono (ceiling configuration), Fluoroscopic x-ray system; Model no. 11328100
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Potential hardware issue for ARTIS icono ceiling system: collision sensor may falsely signal non-existing collisions for up to half an hour when the system is switched off for a longer time (e.g., overnight) and switched on. During this time, system movement is only possible with the "override" function with very slow speed and without further collision protection. This may lead to temporary unavailability of the system for diagnostic and therapeutic usage.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- On October 17, 2023, the firm issued an URGENT MEDICAL DEVICE CORRECTION letter to affected customers. According to the firm's letter, as the unintended behavior is caused by thermal startup effects, its occurrence can be avoided by only switching off the system for a short time. The firm suggests keeping the system running (except during proposed weekly reboot) if availability for treatment of emergency patients is necessary. Please plan for up to half an hour warm-up time after switching on the system. The firm will contact affected customers to modify the mounting of the tube cover to mitigate the probability of occurrence of the issue. You may contact the firm's service organization for an earlier appointment at 1-800-888-7436.
- Quantity in commerce
- 38 worldwide, 11 US
- Distribution
- US Nationwide - Worldwide Distribution: Domestic distribution to FL GA IA IL IN MI MN MO MT PA TX. Foreign distribution to Australia Canada Switzerland China Germany Denmark Egypt Spain Finland United Kingdom Israel Netherlands Sweden South Africa
- Product codes and lots
- UDI-DI 04056869295923 Serial Numbers: All SN's associated with material number 11328100
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0523-2024 | Class II | Ongoing | Voluntary: Firm initiated |