Z-0319-2024Class II

Spectral CT on Rails, Software Version 5.1.0, Model Number 728334

Philips North America LLC / initiated 2023-10-09 / Open, Classified / root cause: software

Philips North America began a recall of Spectral CT on Rails, Software Version 5.1.0, Model Number 728334 on . FDA classified it as Class II and gives the reason as "This recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0319-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
Spectral CT on Rails, Software Version 5.1.0, Model Number 728334
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
This recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This issue could result in the gantry moving to a location the operator does not expect. If this occurs, it presents a risk that the scan could occur at an incorrect location, or that the unintended motion of the gantry could cause a collision with the user or needle. The other part of the recall involves a software crash when the CCT foot pedal is activated (as described in RES ID93363).
Root cause tag
software
FDA root cause
Under Investigation by firm
Action
An URGENT Medical Device Correction notification letter dated 10/9/23 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users Below are short-term precautions to take until the permanent solution is installed. Remain vigilant and confirm gantry position during procedures. Review the Emergency Stop Procedure and Collision Prevention Guidelines provided in Section 6: Working with Specialized Exams of the IFU. Actions planned by Philips to correct the problem Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and install the software update (FCO 72800805) to resolve the issue. Please be assured that maintaining a high level of safety and quality is our highest priority. If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
Quantity in commerce
1 unit
Distribution
US Nationwide distribution in the state of MN.
Product codes and lots
UDI-DI: 00884838103627; Lot Code: 1005

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0319-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0319-2024Class IIOngoingVoluntary: Firm initiated