Spectral CT on Rails, Software Version 5.1.0, Model Number 728334
Philips North America began a recall of Spectral CT on Rails, Software Version 5.1.0, Model Number 728334 on . FDA classified it as Class II and gives the reason as "This recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0319-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- Spectral CT on Rails, Software Version 5.1.0, Model Number 728334
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- This recall is part of a 2-issue recall for the same system. When using the IVC needle position button, on the IVC control box during a series of scans, the plan box may not move to the desired needle position. This issue could result in the gantry moving to a location the operator does not expect. If this occurs, it presents a risk that the scan could occur at an incorrect location, or that the unintended motion of the gantry could cause a collision with the user or needle. The other part of the recall involves a software crash when the CCT foot pedal is activated (as described in RES ID93363).
- Root cause tag
- software
- FDA root cause
- Under Investigation by firm
- Action
- An URGENT Medical Device Correction notification letter dated 10/9/23 was sent to customers. Actions that should be taken by the customer / user in order to prevent risks for patients or users Below are short-term precautions to take until the permanent solution is installed. Remain vigilant and confirm gantry position during procedures. Review the Emergency Stop Procedure and Collision Prevention Guidelines provided in Section 6: Working with Specialized Exams of the IFU. Actions planned by Philips to correct the problem Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and install the software update (FCO 72800805) to resolve the issue. Please be assured that maintaining a high level of safety and quality is our highest priority. If you need any further information or support concerning this issue, please contact the Customer Care Solutions Center (1-800-722-9377).
- Quantity in commerce
- 1 unit
- Distribution
- US Nationwide distribution in the state of MN.
- Product codes and lots
- UDI-DI: 00884838103627; Lot Code: 1005
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0319-2024 | Class II | Ongoing | Voluntary: Firm initiated |