Z-0392-2024Class II

Ingenia 1.5T CX

Philips North America LLC / initiated 2023-10-23 / Open, Classified

Philips North America Llc began a recall of Ingenia 1.5T CX on . FDA classified it as Class II and gives the reason as "Potential component failure in the Gradient Coil could product smoke and/or fire". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0392-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
Ingenia 1.5T CX
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Potential component failure in the Gradient Coil could product smoke and/or fire.
FDA root cause
Under Investigation by firm
Action
The firm mailed URGENT Medical Device Correction Notices dated 10/23/2023 to consignees to notify them of this recall event. Consignees are instructed to identify if they possess any affected devices in their inventory and to follow the provided instructions for when scan aborts occur to prevent the failure mode associated with this recall event. Consignees are to stop the scan immediately and contact Philips Service if any of the failure situations outlined in the notice occur during a scan. Consignees are asked to complete and return the provided response form to Philips no later than 30 days after receipt of the recall notice. Questions can be directed to Customer Care Solutions at 1-800-722-9377.
Quantity in commerce
48 units
Distribution
Domestic: Nationwide Distribution; Foreign: Albania, Andorra, Argentina, Australia, Austria, Bangladesh, Belarus, Belgium, Bolivia, Bosnia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Lebanon, Libya, Luxembourg, Malaysia, Malta, Mexico, Myanmar, Nepal, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Poland, Portugal, Puerto Rico, Romania, Russian Fed., Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, South Korea, Spain, Sudan, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Kingdom, United States, UAE, Venezuela, Vietnam, & Yemen.
Product codes and lots
Product No. 781261 & 781262; UDI-DI: (01)00884838068438(21)77008, (01)00884838068438(21)77016, (01)00884838068438(21)77018, (01)00884838068445(21)77032, (01)00884838068445(21)77034, (01)00884838068445(21)77035, (01)00884838068445(21)77038, (01)00884838068445(21)77040, (01)00884838068445(21)77041, (01)00884838068445(21)77046, (01)00884838068445(21)77049, (01)00884838068445(21)77050, (01)00884838068445(21)77054, (01)00884838068445(21)77056, (01)00884838068445(21)77058, (01)00884838068445(21)77059, (01)00884838068445(21)77061, (01)00884838068445(21)77062, (01)00884838068445(21)77065, (01)00884838068445(21)77066, (01)00884838068445(21)77067, (01)00884838068445(21)77071, (01)00884838068445(21)77072, (01)00884838068445(21)77074, (01)00884838068445(21)77075, (01)00884838068445(21)77078, (01)00884838068445(21)77081, (01)00884838068445(21)77083, (01)00884838068445(21)77084, (01)00884838068445(21)77087, (01)00884838068445(21)77093, (01)00884838068445(21)77094, (01)00884838068445(21)77097, (01)00884838068445(21)77100, (01)00884838068445(21)77101, (01)00884838068445(21)77106, (01)00884838068445(21)77109, (01)00884838068445(21)77110, (01)00884838068445(21)77111, (01)00884838068445(21)77119, (01)00884838068445(21)77120, (01)00884838068445(21)77121, (01)00884838068445(21)77124, (01)00884838068445(21)77126, (01)00884838068445(21)77129, (01)00884838068445(21)77130, (01)00884838068445(21)77131, (01)00884838068445(21)77137; Serial No. 77008, 77016, 77018, 77032, 77034, 77035, 77038, 77040, 77041, 77046, 77049, 77050, 77054, 77056, 77058, 77059, 77061, 77062, 77065, 77066, 77067, 77071, 77072, 77074, 77075, 77078, 77081, 77083, 77084, 77087, 77093, 77094, 77097, 77100, 77101, 77106, 77109, 77110, 77111, 77119, 77120, 77121, 77124, 77126, 77129, 77130, 77131, 77137.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0392-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0389-2024Class IIOngoingVoluntary: Firm initiated
Z-0390-2024Class IIOngoingVoluntary: Firm initiated
Z-0391-2024Class IIOngoingVoluntary: Firm initiated
Z-0392-2024Class IIOngoingVoluntary: Firm initiated
Z-0393-2024Class IIOngoingVoluntary: Firm initiated
Z-0394-2024Class IIOngoingVoluntary: Firm initiated
Z-0395-2024Class IIOngoingVoluntary: Firm initiated