Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0554-2024Class II

Sensica Urine Output System, Catalog Number SCCS1002

C.R. Bard Inc / initiated 2023-11-15 / Open, Classified

C.R. Bard Inc began a recall of Sensica Urine Output System, Catalog Number SCCS1002 on . FDA classified it as Class II and gives the reason as "BD (C.R. Bard) has received user complaints reporting that when the device was connected to WiFi/Internet, the time displayed on the Sensica device changed during a patient monitoring session which resulted in 1)...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0554-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
C.R. Bard Inc
Product
Sensica Urine Output System, Catalog Number SCCS1002
Product code
EXS Urinometer, Electrical
Reason for recall
BD (C.R. Bard) has received user complaints reporting that when the device was connected to WiFi/Internet, the time displayed on the Sensica device changed during a patient monitoring session which resulted in 1) incorrect time displayed on the monitoring screen, 2) incorrect hour block timestamps on the monitoring screen, 3) incorrect hour timestamps on data screens, and 4) incorrect catheter use time.
FDA root cause
Under Investigation by firm
Action
BD (CR Bard) issued an URGENT Medical Device Product Advisory to its consignees on 11/15/2023, via FedEx and email. The notice explained the problem with the device, potential hazard, and requested that the consignee perform Operating System and Sensica Software settings changes per the instructions provided on all devices within your facility s control using the attached directions.
Quantity in commerce
148 units
Distribution
US Nationwide distribution.
Product codes and lots
UDI/DI 00801741215773, Serial numbers: BMGPRMYS001 , BMGPRMYS003, BMGPRMYS004, BMGPRMYS005, BMGPRMYS006, BMGPRMYS007, BMGPRMYS020, BMGPRMYS026, BMGPRMYS028, BMGPRMYS030, BMFWRMSBT4, BMFWRMSJ45, BMFWRMSJ49, BMFZRMS104, BMGNRMS085, BMFSRMSH98, BMFSRMSJ01, BMFVRMSQ62, BMFVRMSQ66, BMFVRMSQ68, BMFYRMSJ41, BMFYRMSJ49, BMFZRMS067, BMFZRMS070, BMFZRMS080, BMFZRMS113, BMFZRMS117, BMGNRMS070, BMFURMSF02, BMFURMSF04, BMFURMSF14, BMFURMSF21, BMFVRMSQ51, BMFVRMSQ52, BMFVRMSQ54, BMFVRMSQ79, BMFVRMSQ83, BMFVRMSQ89, BMFVRMSQ91, BMFWRSJ46, BMGPRMYS011, BMGPRMYS018, BMGPRMYS022 , BMGPRMYS023, BMGPRMYS027, BMFYRMSJ35, BMFYRMSJ37, BMFYRMSJ54, BMFYRMSJ58, BMGNRMS079, BMGNRMSG69, BMGNRMSG74, BMFSRMSH94, BMFSRMSH99, BMFSRMSJ38, BMFSRMSJ44, BMFTRMS138, BMFSRMSJ13, BMFSRMSJ27, BMFSRMSJ33, BMFSRMSJ34, BMFQRMS248, BMFQRMS250, BMFRRMS646, BMFRRMS647, BMGPRMYS008, BMGPRMYS009, BMGPRMYS010, BMGPRMYS013, BMGPRMYS014, BMGPRMYS016, BMFZRMS073, BMFSRMSH78, BMFSRMSH79, BMFSRMSH80, BMFSRMSH81, BMFSRMSH82, BMFSRMSH83, BMFSRMSH93, BMFSRMSJ00, BMFSRMSJ14, BMFSRMSJ15, BMFQRMS246, BMFWRMSJ39, BMFWRMSJ41, BMFSRMSJ36, BMFSRMSJ37, BMFSRMSJ42, BMFTRMS106, BMFTRMS109, BMFTRMS114, BMFTRMS142, BMFURMSF01, BMFURMSF05, BMFSRMSH95, BMFSRMSH97, BMFTRMS120, BMFTRMS122, BMFTRMS123, BMFTRMS135, BMFTRMS136, BMFTRMS137, BMGPRMYS002, BMGPRMYS012, BMGPRMYS019 , BMGPRMYS029, BMFPRMYS009, BMFQRMS244, BMFQRMS272, BMFQRMS276, BMFQRMS286, BMFQRMS293, BMFSRMSJ43, BMFTRMS121, BMGNRMS067, BMGNRMS076, BMFVRMSQ58, BMFVRMSQ70, BMFVRMSQ73, BMFVRMSQ74, BMFVRMSQ90, BMFVRMSR17, BMFWRMSJ27, BMFWRMSJ35, BMFURMSF06, BMFURMSF09, BMFWRMSJ36, BMFWRMSJ51, BMFZRMS068, BMFQRMS294, BMFZRMS105, BMFSRMSJ20, BMFTRMS140, BMFVRMSQ63, BMFVRMSQ64, BMFVRMSQ65, BMFVRMSQ72, BMFVRMSQ75, BMFVRMSQ76, BMFVRMSQ77, BMFVRMSQ78, BMFVRMSQ80, BMFYRMSJ39, BMFZRMS081, BMFZRMS116, BMGNRMS015, BMGNRMS029, BMGNRMS073

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0554-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0554-2024Class IIOngoingVoluntary: Firm initiated