Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with...
Riverpoint Medical, LLC began a recall of Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S (Black/White) REF 3911-714-521 The Iconix... on . FDA classified it as Class II and gives the reason as "Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0518-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Riverpoint Medical, LLC
- Product
- Iconix Knotless 1.4mm Anchor with XBraid S (Blue/Black) REF 3911-714-520 Iconix Knotless 1.4mm Anchor with XBraid S (Black/White) REF 3911-714-521 The Iconix Knotless anchors are intended to be used for soft-tissue to bone fixation in the foot, ankle, knee, hip, hand, wrist, elbow and shoulder.
- Product code
- MBI Fastener, Fixation, Nondegradable, Soft Tissue
- Reason for recall
- Use of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill hole location leading to damage of the suture construct during insertion. Suture damage may lead to suture breaks during use.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- On 10/25/2023, the firm emailed its customer an "URGENT: MEDICAL DEVICE CORRECTION" Letter informing them that the Iconix Knotless Anchor 's Instruction for Use is insufficient to prevent/avoid inserter buckling and insertion failure. The customers were instructed to do the following: - Immediately examine your inventory. If you have the part numbers listed above, notify end user with additional instructions for use. Users may continue to use field inventory once the acknowledgement letter has been completed. - Complete and return the response form as soon as possible to [email protected]. Mark the return shipments with RGA#2081- PDX. - Provide field correction information to any customers to whom you may have distributed or forwarded affected products. You may include a copy of the recall notice with your customer notification. - A copy of the notification letter and of IFU may be posted on or near the affected products, so users are aware of the additional instructions.
- Quantity in commerce
- 999 devices
- Distribution
- U.S. Nationwide distribution in the state of CA.
- Product codes and lots
- Product Code: 3911-7 14-520 UDI-DI Code: 00840277402172 Lot Number: 23080433 Product Code: 3911-7 14-521 UDI-DI Code: 00840277402189 Lot Number: 23081617
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0518-2024 | Class II | Ongoing | Voluntary: Firm initiated |