Z-0507-2024Class II
Stay Safe Cap
Fresenius Medical Care Renal Therapies Group, LLC / initiated 2023-10-10 / Open, Classified / root cause: labeling and UDI
Fresenius Medical Care Holdings, Inc. began a recall of Stay Safe Cap on . FDA classified it as Class II and gives the reason as "Updating the labeling to include additional warnings and contraindications for potential side effects". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0507-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- Stay Safe Cap
- Product code
- KDJ Set, Administration, For Peritoneal Dialysis, Disposable
- Reason for recall
- Updating the labeling to include additional warnings and contraindications for potential side effects.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- Consignees were mailed an IMPORTAND MEDICAL DEVICE INFORMATION notification dated 10/10/23 vis USPS. This notification is to bring awareness to consignees that Fresenius is updating the labeling to their Stay Safe Cap to include additional warnings and contraindications for potential side effects for P/N 05-95012. Increased exposure to povidone-iodine, which is present in the Stay Safe Cap, could lead to iodine-induced hypothyroidism and hypersensitivity. Clinicians are asked to not use these devices in known cases of povidone-iodine allergy. Fresenius is updating labeling to include these medical warnings and contraindications. Any clinical or medical questions can be directed to the firm at 1-855-616-2309. Consignees are asked to return the provided Reply Form by fax to 781-966-9796 or 781-472-7413, ATTN: Quality Department, or by email to [email protected].
- Quantity in commerce
- 18,236,160 eaches
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Model No. 050-95012; UDI-DI: 00840861100835; All Lots.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0507-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.