Z-0522-2024Class II
Medtronic SynchroMed, Model A10
Medtronic, Inc. / initiated 2020-07-13 / Completed / root cause: software
Medtronic Inc. began a recall of Medtronic SynchroMed, Model A10 on . FDA classified it as Class II and gives the reason as "The previous software application version (1.1.300) is missing a decimal separator (a comma) for parameter range guidance values displayed on some of the programming screens: Catheter, Reservoir, Infusion, Bolus, myPTM...". As of , the recall status on record is Completed, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0522-2024
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Medtronic SynchroMed, Model A10
- Product code
- LKK Pump, Infusion, Implanted, Programmable
- Reason for recall
- The previous software application version (1.1.300) is missing a decimal separator (a comma) for parameter range guidance values displayed on some of the programming screens: Catheter, Reservoir, Infusion, Bolus, myPTM, and Alarm.
- Root cause tag
- software
- FDA root cause
- Software Design Change
- Action
- Medtronic issued a SAFETY NOTIFICATION to its consignees on 07/13/2020 by letter. The notice explained the problem with the device and requested the software be upgraded to version 1.1.342.
- Quantity in commerce
- 2543 units
- Distribution
- Foreign Distribution
- Product codes and lots
- GTIN 00643169771031, All serial numbers with software version 1.1.300.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0522-2024 | Class II | Completed | Voluntary: Firm initiated |