Z-0522-2024Class II

Medtronic SynchroMed, Model A10

Medtronic, Inc. / initiated 2020-07-13 / Completed / root cause: software

Medtronic Inc. began a recall of Medtronic SynchroMed, Model A10 on . FDA classified it as Class II and gives the reason as "The previous software application version (1.1.300) is missing a decimal separator (a comma) for parameter range guidance values displayed on some of the programming screens: Catheter, Reservoir, Infusion, Bolus, myPTM...". As of , the recall status on record is Completed, and MDM tags the root cause as software.

Recall record

Recall number
Z-0522-2024
Classification
Class II
Status
Completed
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Medtronic SynchroMed, Model A10
Product code
LKK Pump, Infusion, Implanted, Programmable
Reason for recall
The previous software application version (1.1.300) is missing a decimal separator (a comma) for parameter range guidance values displayed on some of the programming screens: Catheter, Reservoir, Infusion, Bolus, myPTM, and Alarm.
Root cause tag
software
FDA root cause
Software Design Change
Action
Medtronic issued a SAFETY NOTIFICATION to its consignees on 07/13/2020 by letter. The notice explained the problem with the device and requested the software be upgraded to version 1.1.342.
Quantity in commerce
2543 units
Distribution
Foreign Distribution
Product codes and lots
GTIN 00643169771031, All serial numbers with software version 1.1.300.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0522-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0522-2024Class IICompletedVoluntary: Firm initiated