Z-0482-2024Class II

HVAD (HeartWare Ventricular Assist Device) STERILE IMPLANT KIT/IMPLANT KIT OUS, Model Number 1205

HeartWare, Inc. / initiated 2023-10-31 / Open, Classified / root cause: labeling and UDI

Heartware, Inc. began a recall of HVAD (HeartWare Ventricular Assist Device) STERILE IMPLANT KIT/IMPLANT KIT OUS, Model Number 1205 on . FDA classified it as Class II and gives the reason as "Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm may sound and the recommended...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0482-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
HeartWare, Inc.
Product
HVAD (HeartWare Ventricular Assist Device) STERILE IMPLANT KIT/IMPLANT KIT OUS, Model Number 1205
Product code
DSQ Ventricular (Assist) Bypass
Reason for recall
Updates to the HVAD system instructions for use (IFU) and patient manual (PM) will be provided. These updates will clarify (1) the conditions under which an HVAD [Controller Fault] alarm may sound and the recommended troubleshooting actions and (2) instructions relating to the useful life of the HVAD system components
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
Heartware issued an URGENT: MEDICAL DEVICE CORRECTION notice to its consignees on 10/31/2023 via letter (UPS). The notice explained the issue and informed them of the updates to the HVAD IFU and PM (patient manual).
Quantity in commerce
126 units
Distribution
Worldwide
Product codes and lots
Serial numbers: HW1412, HW1413, HW1416, HW1433, HW1435, HW1490, HW1493, HW1495, HW1496, HW1593, HW1599, HW1641, HW1661, HW1662, HW1663, HW1664, HW1695, HW1696, HW1698, HW1699, HW1706, HW1708, HW1709, HW1718, HW1727, HW1730, HW1740, HW1741, HW1747, HW1754, HW1760, HW1764, HW1765, HW1768, HW1773, HW1774, HW1777, HW1778, HW1783, HW1785, HW1796, HW1804, HW1808, HW1810, HW1811, HW1815, HW1819, HW1821, HW1822, HW1823, HW1824, HW1825, HW1826, HW1828, HW1830, HW1831, HW1832, HW1839, HW1841, HW1856, HW1857, HW1876, HW1879, HW1882, HW1884, hw1895, HW1896, HW1897, HW1899, HW1900, HW1901, HW1905, HW1906, HW1907, HW1908, HW1913, hw1914, HW1917, HW1920, HW1921, HW1922, HW1923, HW1924, HW1925, HW1926, HW1927, HW1928, HW1929, HW1930, HW1931, HW1933, HW1934, HW1935, HW1936, HW1937, HW1938, HW1939, HW1940, HW1941, HW1942, HW1943, hw1944, HW1947, HW1950, HW1952, HW1960, HW1961, HW1962, HW1964, HW1965, HW1966, HW1967, HW1968, HW1969, HW1970, HW1971, HW1972, HW1973, HW1974, HW1976, HW1977, HW1978, HW1979, HW1983, HW2023, HW2028

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0482-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0475-2024Class IIOngoingVoluntary: Firm initiated
Z-0476-2024Class IIOngoingVoluntary: Firm initiated
Z-0477-2024Class IIOngoingVoluntary: Firm initiated
Z-0478-2024Class IIOngoingVoluntary: Firm initiated
Z-0479-2024Class IIOngoingVoluntary: Firm initiated
Z-0480-2024Class IIOngoingVoluntary: Firm initiated
Z-0481-2024Class IIOngoingVoluntary: Firm initiated
Z-0482-2024Class IIOngoingVoluntary: Firm initiated
Z-0483-2024Class IIOngoingVoluntary: Firm initiated