Z-0440-2024Class II
Automated Impella Controller (AIC), product number 0042-000-US with software versions V8.4 and V8.4.1
Abiomed, Inc. / initiated 2023-10-10 / Terminated / root cause: software
Abiomed, Inc. began a recall of Automated Impella Controller (AIC), product number 0042-000-US with software versions V8.4 and V8.4.1 on . FDA classified it as Class II and gives the reason as "Pump not detected as connected to controller due to software issue". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0440-2024
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abiomed, Inc.
- Product
- Automated Impella Controller (AIC), product number 0042-000-US with software versions V8.4 and V8.4.1.
- Product code
- OZD Temporary Non-Roller Type Left Heart Support Blood Pump
- Reason for recall
- Pump not detected as connected to controller due to software issue.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- An Urgent Medical Device Correction was issued vail mail October 24. 2023. Customers are advised to examine inventory for product running AIC software version: V8.4 or V8.4.1. Do not to use the product unless software is updated. Contact [email protected] to schedule the software update and return the business reply form. Forward this information to relevant personnel and any facility product may have been transferred to.
- Distribution
- US: AZ, CA, FL, NY & VA.
- Product codes and lots
- UDI-DI (GTIN): 00813502010022, serial numbers IC3772, IC3773, IC5721 and IC4404.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0440-2024 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0441-2024 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY