Z-0557-2024Class II
TRACHEAL INTUBATION FIBERSCOPE OLYMPUS LF-GP, Model Number: LF-GP
OLYMPUS Corporation of the Americas / initiated 2023-09-28 / Open, Classified
Olympus Corporation of the Americas began a recall of TRACHEAL INTUBATION FIBERSCOPE OLYMPUS LF-GP, Model Number: LF-GP on . FDA classified it as Class II and gives the reason as "Fiberscopes do not have adequate data to support that the suction flow rate meets the device specification". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0557-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- TRACHEAL INTUBATION FIBERSCOPE OLYMPUS LF-GP, Model Number: LF-GP
- Product code
- EOQ Bronchoscope (Flexible Or Rigid)
- Reason for recall
- Fiberscopes do not have adequate data to support that the suction flow rate meets the device specification.
- FDA root cause
- Under Investigation by firm
- Action
- An Urgent Medical Device notification dated November 1, 2023 issued via UPS Ground. The notice advises customers to examine inventory for affected serial numbers and quarantine affected product. Contact Olympus Customer Solutions Team (1-800-848-9024, option 3) to arrange for for return. Olympus will inspect and repair if necessary. Customers are to forward notice to any facility affected product may have been transferred to. Olympus requests that you acknowledge receipt of this letter with [email protected].
- Quantity in commerce
- 75
- Distribution
- Worldwide - US Nationwide distribution in the states of AK, AL, AZ, CA, GA, IL, IN, MA, MD, MI, NC, NJ, OH, TX, UT, VA, WA, WV & AP (Armed Forces Pacific) and the countries of OUS: EMEA, Japan and Australia.
- Product codes and lots
- UDI-DI: 04953170340215 & UDI-DI: 04953170061998. Serial numbers: 2515143, 2515181, 2515449, 2515549, 2615760, 2615761, 2615763, 2615825, 2615830, 2615831, 2616057, 2616130, 2616199, 2716417, 2817430, 2917855, 2917867, 2917997, 2918003, 2918010, 2918012, 2918030, 2018337, 2018435, 2018486, 2018543, 2018571, 2018675, 2319768, 2319820, 2510582, 2621001.