Z-0537-2024Class I

AirFit F20 Full Face Mask and User Guide

ResMed Corp / initiated 2023-11-20 / Open, Classified / root cause: design

ResMed Ltd. began a recall of AirFit F20 Full Face Mask and User Guide on . FDA classified it as Class I and gives the reason as "Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0537-2024
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
ResMed Corp
Product
AirFit F20 Full Face Mask and User Guide
Product code
BZD Ventilator, Non-Continuous (Respirator)
Reason for recall
Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical devices/implants are being updated for patients using masks or anyone in close physical contact
Root cause tag
design
FDA root cause
Device Design
Action
On 11/20/23, "Urgent Field Safety Notices" and FAQs were placed on the firms website at www.resmed.com/magnetupdate, and these notices were emailed to healthcare providers, and distributors who were asked to do the following: 1) Provide a copy of this notice and the physician letter to prescribing physicians, and/or any other relevant healthcare professionals, notifying them of the updated labeling (contraindications and warning). 2) Provide a copy of the patient letter to all patients currently using an affected mask with magnets, notifying them of the updated labeling (contraindications and warning). Note, patients are instructed in the patient letter to contact their mask provider for a replacement mask if they are now contraindicated. 3) Provide a replacement mask for contraindicated patients to an alternative mask without magnets in a timely manner. Where an alternative mask is not available, inform patients to consult with their physician. 4) Instruct patients to consult their physician and/or manufacturer of their implant / other medical device if they require additional information on the potential adverse effects of magnetic fields for their particular device (as described in warnings). 5) For alternative mask options, please contact your ResMed or Customer Service representative. 6) Complete and return the customer acknowledgement form via email to [email protected] For any questions, please contact your local ResMed contact or visit www.resmed.com/contact.
Quantity in commerce
11,885,039
Distribution
US nationwide distribution including Puerto Rico, Guam, and Virgin Islands. OUS (foreign) countries of: AE, AR, AT, AU, BB, BD, BE, BG, BH, BM, BR, CA, CH, CL, CN, CO, CW, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FO, FR, GB, GF, GG, GL, GP, GR, GT, GY, HK, HN, HR, HU, ID, IE, IL, IM, IN, IR, IS, IT, JE, JO, JP, KE, KG, KR, KW, LB, LK, LT, LV, LY, MA, MC, MM, MQ, MT, MU, MV, MX, MY, NA, NC, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PM, PT, QA, RE, RO, RS, RU, SA, SD, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UY, VE, VN, WF, YT, ZA
Product codes and lots
All mask lots used with User Guide 638220/version 2020-02, 638238/version 2020-12, 638221/version 2020-01, and prior. UDI-DI/GTIN: 619498634019, 619498634484, 619498634460, 619498634002, 619498634453, 619498634033, 619498634569, 619498634477, 619498634026, 619498634040, 619498634491, 619498634507, 619498634521, 619498634538, 619498640065, 619498640072, 619498634064, 619498640096, 619498634316, 619498640058, 619498634118, 619498640119, 619498634071, 619498634248, 619498634217, 619498634224, 619498634330, 619498634200, 619498634347, 619498634323, 619498634309, 619498634255, 619498634231, 619498634279, 619498634286, 619498634262, 619498634057, 619498634293, 619498634095, 619498634088, 619498634361, 619498634378, 619498634354, 619498640218, 619498640201, 619498640225, 619498634132, 619498634187, 619498634163, 619498634101, 619498634156, 619498634125, 619498634170, 619498640089, 619498640140, 619498634149, 619498640102, 619498640126, 619498640133, 619498634194

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0537-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0536-2024Class IOngoingVoluntary: Firm initiated
Z-0537-2024Class IOngoingVoluntary: Firm initiated
Z-0538-2024Class IOngoingVoluntary: Firm initiated
Z-0539-2024Class IOngoingVoluntary: Firm initiated
Z-0540-2024Class IOngoingVoluntary: Firm initiated
Z-0541-2024Class IOngoingVoluntary: Firm initiated
Z-0542-2024Class IOngoingVoluntary: Firm initiated