Catalog No. 1443 DRESSING CHANGE TRAY / STERILE - LATEX FREE
Busse Hospital Disposables, Inc. began a recall of Catalog No. 1443 DRESSING CHANGE TRAY / STERILE - LATEX FREE on . FDA classified it as Class I and gives the reason as "Firm is recalling all kits containing recalled lots of Nurse Assist Sterile 0.9% Normal Saline. Nurse Assist identified product in distribution for which the required sterility assurance level (SAL) of 10^-6 cannot be...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0593-2024
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Busse Hospital Disposables
- Product
- Catalog No. 1443 DRESSING CHANGE TRAY / STERILE - LATEX FREE
- Product code
- MCY Wound Dressing Kit
- Reason for recall
- Firm is recalling all kits containing recalled lots of Nurse Assist Sterile 0.9% Normal Saline. Nurse Assist identified product in distribution for which the required sterility assurance level (SAL) of 10^-6 cannot be guaranteed. Risks to patients if product is used include blood stream infections, urinary tract infections, or respiratory infections; in the worst-case scenario, there is risk of sepsis or death.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On November 21, 2023, the firm notified affected distributors via URGENT MEDICAL DEVICE RECALL letters. Customers were asked to segregate and destroy any affected kits. Distributors should also inform their customers of the recall. Customers should return any affected product to their distributor. For questions, email [email protected]
- Quantity in commerce
- 536
- Distribution
- US Nationwide distribution.
- Product codes and lots
- UDI-DI 00849233010727 Lots 2210077 2210112 2310275 2310331
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0592-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0593-2024 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-0594-2024 | Class I | Ongoing | Voluntary: Firm initiated |