Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0633-2024Class II

Xstar Kojo Safety Slit Knife, 2.5mm 45 Bevel Up, Part Number 373025

Becton Dickinson & Co / initiated 2023-10-10 / Open, Classified

Beaver Visitec International, Inc. began a recall of Xstar Kojo Safety Slit Knife, 2.5mm 45 Bevel Up, Part Number 373025 on . FDA classified it as Class II and gives the reason as "One lot of Part Number 373025, Xstar Kojo Safety Slit Knife 2.5mm 45 bevel up, incorrectly contains a bevel down blade instead of a bevel up blade". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0633-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Becton Dickinson & Co
Product
Xstar Kojo Safety Slit Knife, 2.5mm 45 Bevel Up, Part Number 373025
Product code
HNN Knife, Ophthalmic
Reason for recall
One lot of Part Number 373025, Xstar Kojo Safety Slit Knife 2.5mm 45 bevel up, incorrectly contains a bevel down blade instead of a bevel up blade.
FDA root cause
Under Investigation by firm
Action
An "URGENT Field Safety Notice" dated 10/10/23 was sent to customers. Actions Required by You Our records show you received one or more affected product. You can recognize the affected product by determining whether the lot number printed on the device label is one of the affected lots. PLEASE perform the following actions: 1. Immediately examine your inventory and determine if the blades on the knives are bevel-up as intended or bevel down. Only destroy the knives in your inventory that have the incorrect bevel orientation (bevel down). If you have further distributed this product, please notify your customers by initiating a sub-recall. Consider all potential users of this product in your user supply chain. You are encouraged to use a copy of this recall notification letter when contacting your customers. 2. Review the enclosed Business Reply Form (BRF) and indicate whether you have the affected product in your possession. 3. Please enter the following information on the enclosed BRF: part number(s), lot number(s) and quantity(ies) of device(s) destroyed. 4. Return the completed BRF and evidence of destruction of affected product to BVI by emailing the attached to [email protected]. BVI values your business and apologizes for the inconvenience this may have caused. Please direct any questions regarding credit to our customer service department by email and include in the subject line: Master Case PIR 00487979. Country Customer Service Email UK [email protected] FR [email protected] DE/ AT [email protected] IT [email protected] ROW [email protected]
Quantity in commerce
2160 units
Distribution
International Distribution to Japan.
Product codes and lots
UDI-DI: 30886158000373; Lot Number: 6049877

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0633-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0633-2024Class IIOngoingVoluntary: Firm initiated