Z-0631-2024Class II

Covidien Valleylab FT10, FT Series Energy Platform, REF: VLFT10GEN, DLVLFT10GEN, VLFT10GENZD

Medtronic, Inc. / initiated 2023-11-16 / Open, Classified / root cause: software

Covidien began a recall of Covidien Valleylab FT10, FT Series Energy Platform, REF: VLFT10GEN, DLVLFT10GEN, VLFT10GENZD on . FDA classified it as Class II and gives the reason as "Due to software issue, New systems may exhibit an error which may prohibit use of new systems resulting in system being inoperable". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0631-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Covidien Valleylab FT10, FT Series Energy Platform, REF: VLFT10GEN, DLVLFT10GEN, VLFT10GENZD
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Due to software issue, New systems may exhibit an error which may prohibit use of new systems resulting in system being inoperable
Root cause tag
software
FDA root cause
Software Manufacturing/Software Deployment
Action
On November 16 and 20, 2023, Covidien LLC (Medtronic Company) issued a "Urgent: Medical Device Correction" 1. Immediately notify all personnel in all care environments in which the Valleylab" FT10 Energy Platform. 2. Update ValleylabTM FT10 Energy Platform to software version 4.0.4. For additional information see customer communication letter 3. Until the software is updated, the Valleylab" FT10 Energy Platform and LigaSure" devices can continue to be used as instructed in the User Guide and per your facility protocols. 4. Complete the attached Customer Confirmation Form and return it as directed to confirm your receipt and understanding of this information.
Quantity in commerce
47,901 units
Distribution
Worldwide distribution - US: Nationwide and surrounding territories.
Product codes and lots
Ref-UDI: VLFT10GEN- 10884521516328, DLVLFT10GEN - 10884521787056 , VLFT10GENZD - 10884521640474/Software: v4.0.1, v4.0.2 and v4.0.3

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0631-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0631-2024Class IIOngoingVoluntary: Firm initiated