Z-0631-2024Class II
Covidien Valleylab FT10, FT Series Energy Platform, REF: VLFT10GEN, DLVLFT10GEN, VLFT10GENZD
Medtronic, Inc. / initiated 2023-11-16 / Open, Classified / root cause: software
Covidien began a recall of Covidien Valleylab FT10, FT Series Energy Platform, REF: VLFT10GEN, DLVLFT10GEN, VLFT10GENZD on . FDA classified it as Class II and gives the reason as "Due to software issue, New systems may exhibit an error which may prohibit use of new systems resulting in system being inoperable". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0631-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Covidien Valleylab FT10, FT Series Energy Platform, REF: VLFT10GEN, DLVLFT10GEN, VLFT10GENZD
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- Due to software issue, New systems may exhibit an error which may prohibit use of new systems resulting in system being inoperable
- Root cause tag
- software
- FDA root cause
- Software Manufacturing/Software Deployment
- Action
- On November 16 and 20, 2023, Covidien LLC (Medtronic Company) issued a "Urgent: Medical Device Correction" 1. Immediately notify all personnel in all care environments in which the Valleylab" FT10 Energy Platform. 2. Update ValleylabTM FT10 Energy Platform to software version 4.0.4. For additional information see customer communication letter 3. Until the software is updated, the Valleylab" FT10 Energy Platform and LigaSure" devices can continue to be used as instructed in the User Guide and per your facility protocols. 4. Complete the attached Customer Confirmation Form and return it as directed to confirm your receipt and understanding of this information.
- Quantity in commerce
- 47,901 units
- Distribution
- Worldwide distribution - US: Nationwide and surrounding territories.
- Product codes and lots
- Ref-UDI: VLFT10GEN- 10884521516328, DLVLFT10GEN - 10884521787056 , VLFT10GENZD - 10884521640474/Software: v4.0.1, v4.0.2 and v4.0.3
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0631-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.