Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0610-2024Class II

LITe Decompression Snake Arm, REF 48080230

Howmedica Osteonics Corp / initiated 2023-11-01 / Open, Classified / root cause: manufacturing process

Stryker Spine began a recall of LITe Decompression Snake Arm, REF 48080230 on . FDA classified it as Class II and gives the reason as "Product arm was manufactured with the incorrect clamp component. As a result, the affected product arm cannot attach to the arm post shaft". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0610-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Howmedica Osteonics Corp
Product
LITe Decompression Snake Arm, REF 48080230
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Product arm was manufactured with the incorrect clamp component. As a result, the affected product arm cannot attach to the arm post shaft.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Stryker issued Urgent Medical Device Correction on 11/20/2023 notifying customers of removal of lot 227764 of the LITe Decompression Tube Snake Arm, catalog number 48080230. A portion of one lot of the LITe Decompression Tube Snake Arm was manufactured with the incorrect clamp component. As a result, the affected Snake Arms cannot attach to the arm post shaft. Customers are to review inventory and remove units from points of use for return to Stryker, share this information with appropriate personnel and return the Business Reply form to [email protected]. If you have any questions or concerns, please contact Regulatory Compliance at (201) 749-8090.
Quantity in commerce
9
Distribution
US: MD, MI, NM, TX & OUS: Canada & France.
Product codes and lots
UDI-DI (GTIN): 07613327263909, lot # 227764.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0610-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0610-2024Class IIOngoingVoluntary: Firm initiated