Z-0993-2024Class II

FilmArray Pneumonia Panel Plus (Pneumoplus), REF: RFIT-ASY-0143, CED 2797, IVD, Rx Only (30 test kit)

BioFire Diagnostics, LLC / initiated 2023-12-14 / Open, Classified

BioFire Diagnostics, LLC began a recall of FilmArray Pneumonia Panel Plus (Pneumoplus), REF: RFIT-ASY-0143, CED 2797, IVD, Rx Only (30 test kit) on . FDA classified it as Class II and gives the reason as "Their is a potential for false positive results when using Pneumonia panel". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-0993-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
BioFire Diagnostics, LLC
Product
FilmArray Pneumonia Panel Plus (Pneumoplus), REF: RFIT-ASY-0143, CED 2797, IVD, Rx Only (30 test kit)
Product code
QDS Mers-Cov And Common Respiratory Pathogens Semi-Quantitative And Quantitative Multiplex Nucleic Acid Detection System
Reason for recall
Their is a potential for false positive results when using Pneumonia panel.
FDA root cause
Under Investigation by firm
Action
On December 14, 2023, Biomerieux, issued a "Urgent Medical Recall" to affected consignees via: FedEx . Biomerieux asked consignees to take the following actions: 1. Distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties that may use this product, including others to whom you may have transferred our product. 2. Use another method to confirm positive results for seasonal Coronavirus from the BIOFIRE Pneumonia Panel or BIOFIRE Pneumonia Panel plus prior to reporting the result (e.g., consider results from testing a nasopharyngeal swab with a different BIOFIRE respiratory panel such as the BIOFIRE Respiratory Panel 2.1). Alternatively, the BIOFIRE Pneumonia Panel or BIOFIRE Pneumonia Panel plus seasonal Coronavirus positive result should not be reported. 3. Complete the Acknowledgement Form in Attachment A (on the following page) and return it to bioMrieux to confirm receipt of this notice. It is important that you return the acknowledgement form to bioMrieux even if you determine that this urgent product correction notice does not impact your facility.
Quantity in commerce
214,080 pouches
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Argentina, Bolivia, Brazil, Chili, China, Colombia, Costa Rica, Dominican Republic, Ecuador, Guatemala, Hong Kong, Honduras, Japan, South Korea, Mexico, Nicaragua, Panama, Peru, Philippines, Paraguay, Singapore, El Salvador, Thailand, Taiwan, Uruguay, Venezuela.
Product codes and lots
UDI: 00815381020314/ All unexpired lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0993-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0993-2024Class IIOngoingVoluntary: Firm initiated
Z-0994-2024Class IIOngoingVoluntary: Firm initiated