Z-0716-2024Class II

Soltive SuperPulsed Laser System TFL Fiber Stripper, Model Numbers TFL-AFS150, TFL-AFS200, TFL-AFS365...

OLYMPUS Corporation of the Americas / initiated 2023-12-04 / Open, Classified / root cause: labeling and UDI

Olympus Corporation of the Americas began a recall of Soltive SuperPulsed Laser System TFL Fiber Stripper, Model Numbers TFL-AFS150, TFL-AFS200, TFL-AFS365, TFL-AFS550, TFL-AFS940 on . FDA classified it as Class II and gives the reason as "The strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0716-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Soltive SuperPulsed Laser System TFL Fiber Stripper, Model Numbers TFL-AFS150, TFL-AFS200, TFL-AFS365, TFL-AFS550, TFL-AFS940
Product code
GEX Powered Laser Surgical Instrument
Reason for recall
The strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a risk of contamination.
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
An URGENT: MEDICAL DEVICE REMOVAL notification letter dated 12/4/23 was sent to customers. Actions to be taken by the end user: Our records indicate that your facility has received one or more affected units. Olympus requests you to take the following actions: 1. Examine your inventory and quarantine any devices with the lot number specified above. 2. Please contact Olympus customer service representative at 1-800-848-9024, option 2, with the quantity and model of affected devices. Olympus will issue a Return Material Authorization to return any affected product at no charge. Olympus will issue a credit to your facility upon return of affected product. 3. Access the Olympus recall portal to indicate that you have received this notification. a. Go to https://olympusamerica.com/recall. b. Enter the recall number "0435 c. Complete the form as instructed. 4. If you have distributed these devices outside your facility, please provide a copy of this letter to those facilities immediately. Olympus requests you to report any complaints, including product labeling issues, to our Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation in addressing this situation. If you require additional information, please do not hesitate to contact me directly by phone at (647) 999-3203 or by e-mail at [email protected].
Quantity in commerce
3001 units
Distribution
US Nationwide. Global Distribution
Product codes and lots
Lot Number: All units sold prior to 2023

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0716-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0716-2024Class IIOngoingVoluntary: Firm initiated
Z-0717-2024Class IIOngoingVoluntary: Firm initiated