Z-0781-2024Class II

Sedecal X Optima URS digital mobile diagnostic x-ray system. A a stationary x-ray unit with a universal...

Sedecal S.A. / initiated 2023-12-13 / Open, Classified / root cause: manufacturing process

SEDECAL SA began a recall of Sedecal X Optima URS digital mobile diagnostic x-ray system. A a stationary x-ray unit with a universal swivel arm. It allows one to take exposures of patients... on . FDA classified it as Class II and gives the reason as "Some parts have fallen off the equipment resulting in a safety risk to patients and/or users". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0781-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Sedecal S.A.
Product
Sedecal X Optima URS digital mobile diagnostic x-ray system. A a stationary x-ray unit with a universal swivel arm. It allows one to take exposures of patients in standing, sitting, or laying position.
Product code
KPR System, X-Ray, Stationary
Reason for recall
Some parts have fallen off the equipment resulting in a safety risk to patients and/or users.
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
On 12/13/2023, the recalling firm emailed the recall letter dated 12/11/2023 to their distributor customers along with attachments for service and maintenance of the units. The letter explained a description of the issue and that the installed devices should be inspected to ensure they are being used in a safe mode (the unit has been properly maintained and does not show any signs of the screws loosening) and to fix any potential issues. This inspection will also be included in all scheduled maintenance tasks to be performed in the future. The letter is to be distributed to all staff involved. The letter included 3 attachments. Annex 1 contains further information regarding the issue. Annex 2 is the acknowledgement of receipt of the notification and understanding of the described problem and its importance. Annex 3 is confirmation that the mandatory field checks or upgrade of the affected unit is complete. The firm reported distributors will provide the information to the medical facility either by using the recalling firm's letter or creating a new letter.
Quantity in commerce
845 units
Distribution
Distribution was made to AL, FL, and NC. There was no military/government distribution.
Product codes and lots
All serial numbers, UDI-DI 08436046002500.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0781-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0781-2024Class IIOngoingVoluntary: Firm initiated