SOLTIVE SuperPulsed Laser Fibers, Model No. TFL-FBX200BS
Olympus Corporation of the Americas began a recall of SOLTIVE SuperPulsed Laser Fibers, Model No. TFL-FBX200BS on . FDA classified it as Class II and gives the reason as "Affected products may potentially have a breached sterile pouch seal. A breached sterile pouch seal may lead to minor procedure delays (if noticed prior to use) or patient infections (if not noticed prior to use in...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0718-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- SOLTIVE SuperPulsed Laser Fibers, Model No. TFL-FBX200BS
- Product code
- GEX Powered Laser Surgical Instrument
- Reason for recall
- Affected products may potentially have a breached sterile pouch seal. A breached sterile pouch seal may lead to minor procedure delays (if noticed prior to use) or patient infections (if not noticed prior to use in patient).
- Root cause tag
- packaging
- FDA root cause
- Under Investigation by firm
- Action
- On December 4, 2023, the firm notified affected customers via Urgent Medical Device Removal letters. Customers were instructed to identify and quarantine the affected model/lot. Affected product should not be used. Olympus will issue a credit to customers upon the return of the affected product. If product was distributed outside the customer's facility, the customer should notify those outside facilities immediately. If you require additional information, please contact the company by phone at (647) 999-3203 or by e-mail at [email protected].
- Quantity in commerce
- 179 units
- Distribution
- Domestic distribution to the following states: AK AR AZ CA FL GA IA ID IN KS KY LA MA MI NC ND NE NJ NY OR PA SC SD TN TX VA WI International distribution to Canada, Latin America, and Europe.
- Product codes and lots
- UDI-DI 00821925043985 Lot KR262848
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0718-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY