Z-0683-2024Class II
Sensis Vibe Hemo
Siemens Medical Solutions USA, Inc. / initiated 2023-11-29 / Open, Classified / root cause: software
Siemens Medical Solutions USA, Inc began a recall of Sensis Vibe Hemo on . FDA classified it as Class II and gives the reason as "The possibility of the Sensis documentation functionality application to crash". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0683-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Sensis Vibe Hemo
- Product code
- DQK Computer, Diagnostic, Programmable
- Reason for recall
- The possibility of the Sensis documentation functionality application to crash.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm issued an URGENT: MEDICAL DEVICE CORRECTION notification dated 11/29/23 by mail to consignees to notify them of this recall event. In the case this failure mode occurs during use, consignees are asked to continue the exam by opening a new study for the same patient without adding any Type 1 reporting events. It is instructed that consignees can compensate this by adding free text comments (procedure notes) only and manually inserting the desired information in those comments. The recalling firm will deploy a software update via Update Instruction AX010/23/S to correct the error. A member of the service organization will contact consignees to schedule the update. Consignees can contact the service organization at 1-800-888-7436 to schedule an earlier appointment. The recall notice is to be disseminated to all users of affected products within consignee organizations and others who may need to be informed. Consignees are to return the provided response form by email to [email protected] within 14 days of receipt.
- Quantity in commerce
- 331 units
- Distribution
- Domestic: CA, CT, & FL.
- Product codes and lots
- Model No. 11007641; UDI-DI 04056869010199; Serial No. 18009, 106245, 18016, 18014, 18011, 18010, 19001, 106244, 18015.