Z-0732-2024Class II

(1) CareEvent, Model No. 866435 (2) CareEvent Upgrade, Model no. 866436

Philips North America LLC / initiated 2023-11-28 / Open, Classified / root cause: software

Philips North America began a recall of (1) CareEvent, Model No. 866435 (2) CareEvent Upgrade, Model no. 866436 on . FDA classified it as Class II and gives the reason as "Push notifications may fail to send to the user under certain conditions. This could potentially result in patient harm due to delay in detection of a change in patient condition". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0732-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
(1) CareEvent, Model No. 866435 (2) CareEvent Upgrade, Model no. 866436
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Push notifications may fail to send to the user under certain conditions. This could potentially result in patient harm due to delay in detection of a change in patient condition.
Root cause tag
software
FDA root cause
Software design
Action
On November 28, 2023, the firm notified affected consignees through URGENT Medical Device Correction letters. This failure is applicable to Apple iOS devices only. If customers use iOS devices (e.g. iPhone, iPad, etc.) to receive event notifications: 1) Please conduct patient monitoring at the central station or patient bedside. 2) Please do not use iOS event notifications for patient monitoring. A Philips representative will contact you to schedule a visit to install a software patch to your device which will correct the current issue. Patch PIC iX 4.2.2 will be provided for PIC iX 4.X and patch CareEvent C.03.07 will be provided for CareEvent C.03.X. If you need any further information, please contact your local Philips representative, or call 1-800-722- 9377.
Quantity in commerce
254 US; 37 OUS
Distribution
Worldwide distribution: US (Nationwide) to states of: CA, GA, IA, MA, MI, MN, NJ, NM, NY, OH, PA, TN, TX, WA, WV; and OUS (Foreign) distribution to countries of: Australia, Austria, Belarus, Belgium, Brazil, Canada, Czech Republic, Denmark, Estonia, Faroe Islands, Finland, France, Germany, India, Indonesia, Ireland, Italy, Japan, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, Norway, Oman, Philippines, Poland, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, United Kingdom, Utd.Arab.Emir., and Uzbekistan
Product codes and lots
UDI-DI: 00884838099128 Software: CareEvent Version C.03.X

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0732-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0731-2024Class IIOngoingVoluntary: Firm initiated
Z-0732-2024Class IIOngoingVoluntary: Firm initiated