Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator, product number DDPA2D1
Medtronic Inc. began a recall of Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator, product number DDPA2D1 on . FDA classified it as Class II and gives the reason as "During internal review of Polaris battery assembly data from the HiPot operation, a test systems engineer noticed that some batteries had gone through the High Potential test twice. This was able to occur because the...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0779-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Cobalt XT DR MRI SureScan Implantable Cardioverter Defibrillator, product number DDPA2D1
- Product code
- LWS Implantable Cardioverter Defibrillator (Non-Crt)
- Reason for recall
- During internal review of Polaris battery assembly data from the HiPot operation, a test systems engineer noticed that some batteries had gone through the High Potential test twice. This was able to occur because the test screen did not prevent retesting after a failure. Some batteries were therefore retested and passed.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Medtronic hand delivered an Urgent Medical Device Recall/Retrieval notice to its sole consignee on 12/07/2023. The notice explained the issue and indicated that the device would be collected and a replacement provided.
- Quantity in commerce
- 1 device
- Distribution
- US Nationwide distribution in the state of Florida.
- Product codes and lots
- UDI/DI 00763000178338, Serial Number RSL604096S