Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0722-2024Class II

MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429

Cook Medical LLC / initiated 2023-12-07 / Open, Classified / root cause: design

Cook Medical Llc began a recall of MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429 on . FDA classified it as Class II and gives the reason as "The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0722-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Cook Medical LLC
Product
MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429
Product code
MQG Accessory, Assisted Reproduction
Reason for recall
The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.
Root cause tag
design
FDA root cause
Device Design
Action
The firm notified affected customers on December 7, 2023 via URGENT Medical Device Correction letters. Per the letter, customers can continue to use their MINC+ device whilst awaiting the defect correction, but Cook advises customers to be vigilant and monitor for device alerts. If the device produces audible and visual alerts to indicate an error with temperature, customers must immediately move any dishes to another incubator. If no other incubators are available, the device can be reset to normal operation by switching off mains power to the device for ten seconds then turning the power back on. The software/firmware update will undergo verification testing including compliance testing prior to release. The updated software/firmware is expected to be available on 01 February 2024. An authorized service agent will contact customers to arrange for the device to be corrected at their facility. The service agent will be updating the software/firmware of the device.
Quantity in commerce
14 US; 101 OUS
Distribution
Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM
Product codes and lots
UDI-DI 00827002444294 Lots AM109365 AM109363 AM109356 AM111934 AM111932 AM111933 AM112041 AM112042 AM112039 AM112040 AM112043 AM112071 AM112072 AM112316

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0722-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0722-2024Class IIOngoingVoluntary: Firm initiated