Z-1005-2024Class II

Guardian 4 Glucose Sensor, REF: MMT-7040A, MMT-7040C1, MMT-7040C2, MMT-7040C3, MMT-7040D1, MMT-7040D2...

Medtronic, Inc. / initiated 2023-11-15 / Open, Classified / root cause: manufacturing process

Medtronic MiniMed, Inc. began a recall of Guardian 4 Glucose Sensor, REF: MMT-7040A, MMT-7040C1, MMT-7040C2, MMT-7040C3, MMT-7040D1, MMT-7040D2, MMT-7040MA, MMT-7040QC1, MMT-7040QC2, MMT-7040QC3 on . FDA classified it as Class II and gives the reason as "Some sensors may have a glucose oxidase (GOX) layer thickness outside of specification (5.5 to 9.2 m). A non-conforming GOX layer thickness could impact sensor function (sensor glucose value) or loss of sensor function...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1005-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
Guardian 4 Glucose Sensor, REF: MMT-7040A, MMT-7040C1, MMT-7040C2, MMT-7040C3, MMT-7040D1, MMT-7040D2, MMT-7040MA, MMT-7040QC1, MMT-7040QC2, MMT-7040QC3
Product code
OZP Automated Insulin Dosing Device System, Single Hormonal Control
Reason for recall
Some sensors may have a glucose oxidase (GOX) layer thickness outside of specification (5.5 to 9.2 m). A non-conforming GOX layer thickness could impact sensor function (sensor glucose value) or loss of sensor function due to the sensor layer being too thin. The potential impact resulting from this issue ranges from user inconvenience to over/under delivery of insulin, possibly resulting in hyperglycemia or hypoglycemia, if the sensor glucose values are affected. The issue was identified by the manufacturing team during the execution of Filmetrics Measurement for a Guardian Sensor lot.
Root cause tag
manufacturing process
FDA root cause
Process change control
Action
On 11/15/23, recall notices were mailed or emailed to patients, distributors, and healthcare professionals who were told the following: 1) If you have affected devices discard and contact the firm to receive replacement sensors in the following ways: by completing the web form at medtronicdiabetes.com/G4S-form, contacting 24-Hour Technical Support at 800-646-4633, option 1, or email [email protected]. 2) If you cannot identify the LOT number of the sensor you are currently wearing, remove the sensor and replace your sensor with a non-impacted sensor. 3) Please complete and return the confirmation form via email to [email protected] Distributors were asked to either 1) notify impacted customers using the Correction Letter and provide status updates to the recalling firm regarding customer responses, or 2) provide the recalling firm with customer addresses so they can notify impacted customers. If you have any questions, please contact the firm at 800-646-4633, option 1.
Quantity in commerce
114,758
Distribution
Worldwide - US Nationwide distribution including in the states of PR, AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Austria, Belgium, Canary Islands, Czech Republic, Denmark, Finland, France, Germany, Greece, Iceland, India, Ireland, Israel, Italy, Kazakhstan, Kuwait, Liechtenstein, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Saudi Arabia, Slovenia, Spain, Sweden, Switzerland, United Kingdom.
Product codes and lots
REF/UDI-DI/Lot: MMT-7040A/20763000413389/HG72B2C, HG72CQQ, HG73VB7, HG73VC8, HG73VFT, HG73WFR, HG7441E, HG745F9, HG7486X, HG74AJ1, HG74SUR, HG74T9Q, HG74U36, HG78LMS, HG79Y9U; MMT-7040C1/00763000555276, 00763000413392, 20763000413396, 20763000555270, 20763000712475, 00763000712471/F1323, F1923, G0323, G1723, G2023, G2623, G2923, HG6RV7S, HG6TGNU, HG6TJ2Y, HG6TW97, HG6TWVU, HG6UNF2, HG6V8NN, HG6VB8S, HG6VNX3, HG6WNE1, HG6XFPR, HG6XH54, HG6XQCG, HG6XZJX, HG6Y0W1, HG6Y3QN, HG6YEK6, HG6YFAF, HG6YFHL, HG6YJAR, HG6YJBD, HG6YNGD, HG6Z7N5, HG6ZARN, HG7038Y, HG70CW5, HG70DAM, HG70JSU, HG70LRC, HG70QWA, HG71VMZ, HG72DB5, HG72G9C, HG72NV9, HG72R1P, HG72RBD, HG72SE3, HG72TPT, HG72TRV, HG72V3J, HG72ZFP, HG7340A, HG73EJX, HG73RNR, HG745SV, HG747E3, HG7480Q, HG749N6, HG74F93, HG75TSX, HG75XA1, HG76HFH, HG76RYY; MMT-7040C2/20763000555287/HG6RC39, HG6RXB5, HG6T4T7, HG6T8RV, HG6TJ2Y, HG6WKZU, HG6WLKS, HG6XFPR, HG6XH54, HG6XLSD, HG6XNH3, HG6Y5C6, HG6Z1X5, HG713AK, HG73EGX, HG73EJX, HG73FJP, HG73Q24, HG73QPM, HG73RNR, HG7619X, HG764J9, HG76CWC, HG77FMK; MMT-7040C3/20763000555294/HG6XNH3, HG73RNR; MMT-7040D1/00763000413484/HG6THY1, HG6V7VV; MMT-7040D2/00763000413491/HG6TL44; MMT-7040MA/20763000519692/HG74460; MMT-7040QC1/00763000555313, 20763000555317/HG6WT9A, HG6WUGE, HG73PH3; MMT-7040QC2/20763000555324, 763000555320/HG6WT9A; MMT-7040QC3/20763000555331/HG6WT9A

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1005-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1005-2024Class IIOngoingVoluntary: Firm initiated