Z-0778-2024Class II
Accure Laser System, Model: PFMS00004
Accure Acne, Inc / initiated 2023-12-06 / Open, Classified
Accure Acne, Inc began a recall of Accure Laser System, Model: PFMS00004 on . FDA classified it as Class II and gives the reason as "Laser system's spot tracking system may become misaligned resulting in the possible risk of delivery of higher-than-expected laser energy to epidermal tissue, which may cause blistering". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0778-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Accure Acne, Inc
- Product
- Accure Laser System, Model: PFMS00004
- Product code
- GEX Powered Laser Surgical Instrument
- Reason for recall
- Laser system's spot tracking system may become misaligned resulting in the possible risk of delivery of higher-than-expected laser energy to epidermal tissue, which may cause blistering.
- FDA root cause
- Use error
- Action
- On 12/6/23, "Field Safety Alert(s)" were emailed to customers informing them of the following: 1) Following these steps will minimize the likelihood of Region of Interest (ROI) misalignment: Make sure that during treatment that the handpiece tip movement is minimized and to not lift off the tip until the emission is complete and foot pedal released. Remember this order of operations to reduce the potential of misalignment: "Tip down "Pedal down "Emission completes "Pedal up "Tip up 2) Monitor the patient during treatment for unexpected levels of pain. If higher than usual stop treatment and contact the recalling firm. 3) Immediately discontinue therapy if any Skin Overtemperature alarms are observed while treating with the device in Boost mode. This could be an indication that the ROI is misaligned. Contact your Accure representative before any further use of the device. 4) Immediately discontinue therapy if any PETs are observed that are above the green range displayed on the device touchscreen. This could be an indication that the ROI is misaligned. Contact your recalling firm service representative. 5) Ask your patients to contact you immediately if they experience prolonged erythema (longer than 2 hours) or blistering. Blistering is a potential residual risk and is not expected. If it occurs, it may require medical intervention. It could also be an indication that the ROI is misaligned. 6) If ROI misalignment is suspected suspend all further use of the laser and contact the recalling firm. The firm is releasing a software upgrade to maintain the ROI in the intended monitored location regardless of movement of the handpiece tip, and will safely detect and disable the system in the event that the ROI has become misaligned. Acknowledge, and send your response to the firm, see Field Safety Notice. Customers with questions can contact the firm via email at [email protected]
- Quantity in commerce
- 22
- Distribution
- US distribution to states of: CO, CA, CT, NY, NJ, MN, LA, OH, MA, TX
- Product codes and lots
- UDI-DI: 08059173392230, Serial Numbers: LAD0000-0004, LAD0000-0014, PROT01-0720, LAD0000-0012, LAD0000-0005, LAD2216-0623, LAD2356-0723, LAD0000-0003, PROTO07-1120, LAD2076-0623, LAD0000-0011, LAD0000-0008, LAD0000-0010, LAD2357-0723, LAD2075-0623, LAD2232-0723, LAD0000-0013, LAD0000-0009, LAD0000-0006, LAD0000-0001, PROT06-1120, LAD2507-0723
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0778-2024 | Class II | Ongoing | Voluntary: Firm initiated |