Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR NEURO PACK, kit number...
American Contract Systems, Inc. began a recall of Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR NEURO PACK, kit number AGNE20J; b) ADULT CRANIOTOMY PACK, kit number... on . FDA classified it as Class II and gives the reason as "During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked"...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0827-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- American Contract Systems, Inc.
- Product
- Custom procedural convenience kits and trays, general hospital use, labeled as: a) IR NEURO PACK, kit number AGNE20J; b) ADULT CRANIOTOMY PACK, kit number LLCR99; c) LAMINECTOMY PACK, kit number LLLC26; d) KIT, CRANIOTOMY, kit number MMCR11J; e) KIT, NEURO SPINE, kit number MMNS20H; f) IR NEURO PACK, kit number OSIN77; g) NEURO SAH, kit number SANE32H; h) NEURO SAH, kit number SANE32I; i) S78FCH CRANI PACK (PS 132432), kit number UICR10U; j) CRANI PACK (PS039498), kit number UICR33AW; k) Laminectomy, kit number UILM77AR; l) DIAGNOSTIC NEURO IR TRAY, kit number WENE21D; m) DIAGNOSTIC NEURO IR TRAY, kit number WENE21F; n) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H; o) DIAGNOSTIC NEURO IR TRAY, kit number WENE21H-01; p) NEURO IR TRAY, kit number WENI56F; q) NEURO IR TRAY, kit number WENI56H; r) NEURO IR TRAY, kit number WENI56H-01
- Product code
- OJG Neurological Tray
- Reason for recall
- During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Owens & Minor issued an URGENT: MEDICAL DEVICE CORRECTION notice to it consignees on 11/29/2023 via email. Then notice explained the problem, risk to health, and requested the following actions be taken: If you received any of the batches/lots identified, please segregate and quarantine all affected product. Add warning labels to all kits/cases impacted: A labeling template is enclosed for printing warning labels which should be affixed to all affected product in inventory. This label should be affixed to each affected kit and case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. " Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. " Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified of this field correction and maintain records of effectiveness. Report any quality problems or adverse events associated with the products listed in this notification to [email protected].
- Quantity in commerce
- 2717 units
- Distribution
- US Nationwide distribution.
- Product codes and lots
- a) AGNE20J, UDI/DI 00191072183187, Lot Numbers: 934231, exp. 10/30/2023; 900231, exp. 4/10/2024; 990231, exp. 1/10/2024; 687221, exp. 11/9/2023; 720221, exp. 10/7/2023; 757221, exp. 8/31/2023; 870221, exp. 5/10/2023; 702221, exp. 10/25/2023; 795221, exp. 7/24/2023; b) LLCR99, UDI/DI 00191072162755, Lot Numbers: 756221, exp. 9/1/2023; 720221, exp. 10/7/2023; 654221, exp. 12/12/2023; 949231, exp. 2/20/2024; c) LLLC26, UDI/DI 00191072171900, Lot Numbers: 749221, exp. 9/8/2023; 695221, exp. 11/1/2023; 997231, exp. 1/3/2024; 920231, exp. 3/21/2024; d) MMCR11J, UDI/DI none, Lot Numbers: 744221, exp. 9/13/2023; 939231, exp. 3/2/2024; 876231, exp. 5/4/2024; e) MMNS20H, UDI/DI none, Lot Numbers: 963231, exp. 2/6/2024; 959231, exp. 2/10/2024; f) OSIN77, UDI/DI 00191072170705, Lot Numbers: 780221, exp. 8/8/2023; 728221, exp. 9/29/2023; 668221, exp. 11/28/2023; 991231, exp. 1/9/2024, 960231, exp. 2/9/2024; 939231, exp. 3/2/2024; UDI/DI 00191072190031, Lot Numbers: 912231, exp. 3/29/2024; 903231, exp. 4/7/2024; g) SANE32H, UDI/DI none, Lot Numbers: 838221, exp. 6/11/2023; 889221, exp. 4/21/2023; h) SANE32I, UDI/DI none, Lot Numbers: 673221, exp. 11/23/2023; 730221, exp. 9/27/2023; 784221, exp. 8/4/2023; 911231, exp. 3/30/2024; 969231, exp. 1/31/2024; i) UICR10U, UDI/DI none, Lot Numbers: 990231, exp. 1/10/2024; 921231, exp. 3/20/2024; 885231, exp. 4/25/2024; j) UICR33AW, UDI/DI none, Lot Numbers: 658221, exp. 12/8/2023; 931231, exp. 3/10/2024; 876231, exp. 5/4/2024; k) UILM77AR, UDI/DI 00191072139184, Lot Numbers: 953231, exp. 2/16/2024; 905231, exp. 4/5/2024; l) WENE21D, UDI/DI 00191072167958, Lot Numbers: 839221, exp. 6/10/2023; m) WENE21F, UDI/DI 00191072167958, Lot Numbers: 819221, exp. 6/30/2023; n) WENE21H, UDI/DI 00191072179326, Lot Numbers: 729221, exp. 9/28/2023; 686221, exp. 11/10/2023; o) WENE21H-01, UDI/DI 00191072179326, Lot Numbers: 980231, exp. 1/20/2024; p) WENI56F, UDI/DI 00191072169341, Lot Numbers: 799221, exp. 7/20/2023; q) WENI56H, UDI/DI 00191072179333, Lot Numbers: 700221, exp. 10/27/2023; r) WENI56H-01, UDI/DI 00191072179333, Lot Numbers: 652221, exp. 12/14/2023; 996231, exp. 1/4/2024
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0816-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0817-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0818-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0819-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0820-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0821-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0822-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0823-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0824-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0825-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0826-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0827-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0828-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0829-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0830-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0831-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0832-2024 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0833-2024 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.