TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section...
St. Jude Medical, Atrial Fibrillation Division, Inc. began a recall of TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section, Sterile on . FDA classified it as Class II and gives the reason as "When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model. This results in some...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-0814-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott Medical
- Product
- TactiFlex Ablation Sensor Enabled Ablation Catheter, Models A-TFSE-FJ, 7.5 F shaft and an 8F distal section, Sterile.
- Product code
- OAE Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Fibrillation
- Reason for recall
- When the catheters are used with the EnSiteX EP System, the system does not correctly recognize the catheter as the appropriate product model and instead reads and displays it as a different model. This results in some features specific to the catheter do not appear on the display and the catheter image may also appear inverted.
- FDA root cause
- Under Investigation by firm
- Action
- The recalling firm issued letters dated 12/18/2023 via hand-delivery and FedEx beginning 12/18/2023. The letter explained the issue, the scope of the problem, the impact and associated risks, and next steps to help reduce risk. These steps were: (1) Do not use any remaining inventory from the affected lots listed in Appendix A; (2) Complete and return the accompanying Acknowledgment Form; (3) Return all remaining unused devices from the affected lots; and (4) Forward the letter to anyone within their organization who may need to be notified. Replacements of returned affected devices with a similar curve are available and the customer's Abbott representative can assist in returning these devices and obtaining replacements.
- Quantity in commerce
- 483 units
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CO, DC, DE, FL, GA, IA, IL, KS, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NV, NY, OH, PA, SC, TN, TX, UT, VA, WA, and WV. The country of Canada.
- Product codes and lots
- Lot numbers 10073391, 10075618, 10080798, 10086492, 10087953, 10089624, 10092157, 10093066, and 10095363, Exp. 11/30/2024, GTIN 05415067034618.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0814-2024 | Class II | Ongoing | Voluntary: Firm initiated |