Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. began a recall of Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12) on . FDA classified it as Class II and gives the reason as "Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0861-2024
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips Medical Systems Nederland B.V.
- Product
- Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.
- Root cause tag
- software
- FDA root cause
- Under Investigation by firm
- Action
- The recalling firm issued an URGENT: MEDICAL DEVICE CORRECTION notification dated 12/08/2023 by certified mail to consignees. Consignees with affected devices are asked to utilize the provided workaround instructions. Additionally, consignees are asked to make all users aware of this recall notification and keep a copy of this notification with system documentation. Getinge Field Service Engineers will be dispatched to the field to install software updates to affected devices. Consignees with any questions can contact the recalling firm by phone at 1-800-722-9377. Consignees are asked to return the provided response form to the recalling firm at [email protected].
- Quantity in commerce
- 108 systems (14 US, 1 Canada, 93 ROW)
- Distribution
- Worldwide - US Nationwide distribution in the states of CO, FL, IL, MS, NV and TX. The country of Canada.
- Product codes and lots
- (1) System Code: 722079; (1) UDI/DI: 00884838085268; (1) Serial Number: 1005, 866, 1361, 952, 1241, 1169, 756, 703351, 1135, 1122, 824, 748, 630, 638, 703385, 812, 703346, 1303, 1313, 893, 768, 929, 703638, 1357, 703810, 934, 1202, 703461, 703706, 1172, 703959, 703845, 703705; (2) System Code: 722224; (2) UDI/DI: 00884838099258; (2) Serial Number: 539, 395, 278, 134, 730, 252, 370, 36, 572, 645, 272, 495, 781, 62, 1388, 352, 1655, 690, 555, 895, 2165, 861, 60, 20, 648, 2231, 1778, 1431, 778, 477, 917, 1718, 388, 2042, 150, 764, 2036, 1421, 157, 879, 212, 144, 1463, 347, 575, 2187, 355, 2576, 1332, 424, 387, 863, 211, 474, 712, 549, 2309, 799, 2283, 639, 1700, 2071, 537, 1589, 2208, 2057, 771, 772, 1216, 1667;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0861-2024 | Class II | Ongoing | Voluntary: Firm initiated |