Z-0974-2024Class II

Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient...

Philips North America LLC / initiated 2022-01-06 / Open, Classified / root cause: design

Philips North America began a recall of Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system on . FDA classified it as Class II and gives the reason as "Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0974-2024
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
Root cause tag
design
FDA root cause
Device Design
Action
Beginning in January 2022, Philips Field Service Engineers (FSE) visited impacted customer sites to replace the magnetic castor wheels of the HA Flextrack Trolley II with non-magnetic castor wheels. All reachable systems were addressed as of August 10, 2023.
Quantity in commerce
2
Distribution
Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.
Product codes and lots
UDI-DI 00884838088115 Ingenia Serial number / Accessory Serial Number 45294 100029 45294 100030

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0974-2024
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0973-2024Class IIOngoingVoluntary: Firm initiated
Z-0974-2024Class IIOngoingVoluntary: Firm initiated
Z-0975-2024Class IIOngoingVoluntary: Firm initiated